Hertel exophthalmometry measures the distance from the lateral orbital rim to the corneal apex, in millimetres.
Indian studies give mean values of about 14.5–15.5 mm, with normal readings up to about 21–22 mm.
A difference of more than 2 mm between the eyes is significant, whatever the absolute values.
Always record the base (bar) reading so that follow-up measurements use the same setting.
The Hertel cannot be used reliably when the lateral orbital rim is fractured or displaced.
Measuring proptosis is quick, cheap and essential in thyroid eye disease, orbital tumours and trauma. This page explains how to use the Hertel exophthalmometer, what normal values look like in Indian adults, and the common sources of error.
How to measure
Sit facing the patient at the same eye level. Ask the patient to look straight ahead, at your eye or a distant target.
Place the instrument's footplates on the deepest part of each lateral orbital rim, firmly but gently.
Read the base value on the bar and write it down.
Close your left eye and read the patient's right eye with your right eye: line up the two marks (or the red line) in the mirror to avoid parallax, then read where the corneal apex sits on the scale.
Repeat for the other eye with your other eye.
Record as, for example: "Hertel 17 / 21 mm at base 102".
Normal values
Population
Mean (mm)
Range or upper limit (mm)
North India (1,000 people aged over 5 years)
14.94 ± 2.43; men 15.4, women 14.5
8–22
South India (387 people aged 7–70 years)
About 14.7 (right eye)
Highest in the 3rd to 5th decades
India, tertiary hospital (1,000 people aged 5–90 years)
Not stated
8–21
Thailand
Men 15.0, women 13.6
Mean difference between eyes 0.35
China (Han, Shenyang)
15.0
Upper limit about 19.3 in adults
Values rise through childhood, peak in middle age and fall a little in older people. Men read about 1 mm higher than women.
Normal values differ by ethnicity; values quoted from Western textbooks (often up to 20–21 mm) are close to Indian upper limits but not identical.
Compare the two eyes and compare with earlier readings: a change over time or an asymmetry matters more than a single number.
When a reading misleads
Looks like proptosis but is not
Looks normal but there is proptosis
High axial myopia (large globe)
Bilateral, symmetrical proptosis within the normal range
Lid retraction, as in thyroid eye disease, which exposes more sclera
Proptosis with a displaced lateral rim after a fracture
Enophthalmos of the other eye (blow-out fracture, silent sinus syndrome)
Different base value from the last visit
Facial asymmetry or a shallow orbit
Footplates placed on soft tissue instead of bone
Other ways to measure
Luedde exophthalmometer: a transparent ruler held on one lateral rim; simple and cheap, measures one eye at a time.
Naugle exophthalmometer: rests on the upper and lower orbital rims, so it works when the lateral rim is fractured.
CT or MRI: measures globe position relative to the interzygomatic line and shows the cause.
Clinical view from above: looking down from behind the seated patient shows asymmetry quickly but does not give a number.
Where it is used
Thyroid eye disease: a change of 2 mm or more is part of activity and severity scoring and of judging response to treatment. See the viva hot seat station on thyroid eye disease.
Orbital tumours and inflammation: baseline and follow-up of axial proptosis.
Orbital fractures: enophthalmos of 2 mm or more is one indication considered for repair.
Where residents slip
Not recording the base value, so the next reading cannot be compared.
Reading with the wrong eye and introducing parallax error.
Calling a high myope proptosed.
Pressing the footplates so hard that the patient pulls back.