Short answer: Goldmann tonometry assumes an average corneal thickness, so thin corneas read falsely low and thick corneas falsely high. However, there is no reliable formula to "correct" IOP for central corneal thickness (CCT). Use CCT as a risk factor (thin CCT increased glaucoma risk in OHTS) rather than as a correction, and remember that corneal biomechanics also matter.
| Cornea | Goldmann reading | Example |
|---|---|---|
| Thin | Underestimates true IOP | Thin corneas, after myopic LASIK or PRK |
| Thick (healthy) | Overestimates true IOP | Some "ocular hypertension" patients with thick corneas |
| Thick but oedematous | Often underestimates | Corneal oedema makes the cornea softer despite being thicker |
No validated formula is accurate enough for individual patients. Use CCT as a risk factor and interpret IOP readings with it in mind, rather than applying a correction.
Yes. After myopic laser refractive surgery the cornea is thinner and Goldmann tonometry tends to underestimate IOP. Record pre-operative IOP and CCT if possible.
Thin corneas lead to IOP being underestimated, and thin CCT was also an independent risk factor for developing glaucoma in OHTS.