Slit-Lamp Examination: Systematic Technique, Illumination Methods and Grading Scales

For: PG residents

Last updated: November 2026

Key points in 60 seconds

  • Set the eyepieces and interpupillary distance before every clinic, and line the patient's lateral canthus up with the mark on the headrest.
  • Work in the same order every time: lids, lashes, tear film, conjunctiva, cornea, anterior chamber, iris, lens, anterior vitreous.
  • Start with diffuse light at low magnification, then narrow the beam and raise magnification for detail.
  • Grade what can be graded (van Herick, cells, flare, staining) so that the next examiner can compare.
  • Do not finish without looking at the anterior vitreous for cells and pigment.

The slit lamp is the most-used instrument in the eye OPD, and most missed findings come from an unsystematic examination rather than a poor instrument. This page sets out a routine, matches each illumination method to the findings it shows best, and gathers the grading scales you need to record what you see.

Setting up

  1. Focus each eyepiece for your own eyes (turn fully plus, then back until the focusing rod is sharp) and set the interpupillary distance.
  2. Seat the patient with the chin in the cup and forehead against the band; raise the chin rest until the lateral canthus is level with the canthal mark.
  3. Clean the chin rest and forehead band between patients.
  4. Start at low magnification (about 10×) with a wide, diffused beam and moderate brightness, then increase as needed.

A routine that misses nothing

StructureLook for
Lids and lashesMargin position, meibomian gland orifices and expression, crusts, collarettes, misdirected lashes, lumps
Tear filmMeniscus height, debris, break-up time with fluorescein
ConjunctivaInjection pattern, follicles or papillae (evert the upper lid), chemosis, staining
CorneaEpithelium, stroma and endothelium in turn; staining pattern, infiltrates, keratic precipitates, guttae
Anterior chamberDepth centrally and peripherally (van Herick), cells, flare, hypopyon, hyphaema
Iris and pupilNodules, transillumination, neovascularisation at the pupil margin, synechiae, pupil reactions
LensOpacities by layer, pseudoexfoliation material, phacodonesis; after dilation, retroillumination
Anterior vitreousCells, pigment (Shafer's sign), haemorrhage

Illumination methods and what each shows

MethodHowUse it for
DiffuseWide beam with the diffuser, low magnificationOverview of lids, conjunctiva and cornea; photographs
Direct focal: optic sectionVery narrow beam at 30–60°, focused on the tissueDepth of a corneal lesion, corneal thickness, lens layers, van Herick
Direct focal: parallelepipedBeam 1–2 mm wideArea and texture of corneal lesions, scars, infiltrates
Conical beamShort, narrow beam in a dark room at high magnificationAnterior chamber cells and flare
IndirectBeam focused beside the area of interestEpithelial microcysts, oedema, subtle infiltrates
RetroilluminationLight reflected back from the iris or fundusCorneal vessels and keratic precipitates; iris transillumination; lens vacuoles and posterior capsule opacity
Specular reflectionAngle of illumination equals the angle of viewing, high magnificationEndothelial mosaic, guttae, tear film lipid layer
Sclerotic scatterBeam on the limbus; light travels through the cornea by total internal reflectionFaint central corneal haze or oedema
Cobalt blue with yellow barrier filterFluorescein instilledStaining, break-up time, Seidel test, RGP fit
Red-free (green)Green filterVessels, haemorrhages, rose bengal or lissamine green staining

A shorter table aimed at optometrists, with keratometry, is in Keratometry and Slit-Lamp Illumination Techniques.

Measuring with the slit beam

  • Most slit lamps show the beam height (and often its width) in millimetres. Shorten the beam until it just spans a lesion to record its size, for example an ulcer of 3.0 × 2.5 mm.
  • Record the height of a hypopyon or hyphaema in millimetres in the same way, so a change is obvious at the next visit.
  • Note the size of the epithelial defect (fluorescein) and the infiltrate (white light) separately in keratitis.

Grading scales worth knowing

Van Herick (peripheral anterior chamber depth at the temporal limbus, optic section at 60°)

GradePeripheral chamber depth vs corneal thicknessMeaning
4Equal to or more than the corneal thicknessWide open
31/4 to 1/2Open
21/4Narrow: do gonioscopy before dilating
1Less than 1/4Very narrow: gonioscopy needed
0Iris touching corneaClosed

Anterior chamber cells (SUN: cells in a 1 mm × 1 mm beam)

Grade00.5+1+2+3+4+
Cells in field<11–56–1516–2526–50>50

Flare (SUN): 0 none; 1+ faint; 2+ moderate, iris details clear; 3+ marked, iris details hazy; 4+ intense, fibrin or plastic aqueous. More in the uveitis work-up and gonioscopy grading pages.

The Seidel test

Paint concentrated fluorescein (a moistened strip) over the suspected leak and watch in cobalt blue light. A dark stream of diluted fluorescein flowing through the bright orange-green dye shows an aqueous leak. Gentle pressure on the globe can bring out a slow leak. Use it after trauma, surgery (wound, bleb) and in corneal melts.

Posterior segment and angle at the slit lamp

With a 90D, 78D or wide-field lens the slit lamp also examines the disc, macula and periphery, and with a gonio lens the angle. Lens choice is covered in Ophthalmic Examination Lenses Compared.

Recording and photographs

  • Draw what you see with labels and sizes; a quick sketch is often more useful than a long description.
  • A phone adapter on the eyepiece gives usable photographs for follow-up. Use diffuse light for the overview and a narrow beam for depth, and record which eye and which view.
  • Obtain consent for images, store them in the patient record, and never share identifiable images outside clinical care.

🛒 Where to buy slit-lamp accessories

These are Amazon India search links and affiliate links: Sadbhaav Eye Clinic may earn a small commission from qualifying purchases at no extra cost to you. Check the seller, warranty and genuine-product guarantee before you buy; for costly devices, compare with authorised dealers.

Where residents slip

  • Going straight to high magnification and missing the overall picture, such as a lid lesion or ptosis.
  • Not everting the upper lid in a red eye or foreign body sensation.
  • Calling cells without a dark room and a proper conical beam.
  • Forgetting to look at the anterior vitreous.
  • Dilating a hyperope with a grade 1–2 van Herick without gonioscopy.

Related: Keratometry and slit-lamp techniques (optometrists), Uveitis work-up, Gonioscopy grading, Anterior segment grading scales.

Revise with flip cards

Key sources

  • Standardization of Uveitis Nomenclature (SUN) Working Group. Am J Ophthalmol 2005 (cell and flare grading)
  • Van Herick W, Shaffer RN, Schwartz A. Estimation of width of angle of anterior chamber. Am J Ophthalmol 1969
  • EyeWiki: Ophthalmic Exam Lenses

📘 Want the exam-ready deep dive? This topic, and much more is covered in Ophthalmology Explorer (A High-Yield Clinical Reference for Ophthalmology Residents, Fellows & Board Exams) — available on Kindle. See all my books.

How to cite this page

Patel D. Slit-Lamp Examination: Systematic Technique, Illumination Methods and Grading Scales. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/slit-lamp-examination-technique.html. Accessed [date].

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy

Consultant, Cataract & Refractive Surgery — among a small number of AIIMS-trained ophthalmologists practising in Gujarat. Read full credentials & experience or view his 28 publications on ResearchGate.

📚 Also the author of OKAT Decoded, FRCOphth Decoded, FICO Decoded, OKAP & ABO Decoded, FEBO Decoded and other high-yield ophthalmology exam-prep books used by residents and PG aspirants.

Want to know more about treatment? Read about the Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.