Key points in 60 seconds
RPC Rounds for Optometrists · B.Optom study guide
Keratometry and the slit lamp are the two instruments an optometrist uses most after the phoropter. Keratometry gives corneal curvature for contact lenses and IOL calculations; the slit lamp, used with the right illumination, shows everything from contact lens fit to early keratoconus.
| Instrument | Design | Notes |
|---|---|---|
| Javal-Schiøtz | Fixed image size, variable object size; one-position (measures one meridian at a time) | Mires are a red step and a green rectangle; overlap of steps gives astigmatism |
| Bausch and Lomb (Helmholtz type) | Variable image size, fixed object; two-position (both meridians without rotating) | Plus and minus signs on circles; align the pluses for horizontal and minuses for vertical |
| Automated keratometer | Measures reflected light points at several radii | Fast; included in autorefractors and biometers |
| Topographer and tomographer | Placido or Scheimpflug imaging of the whole cornea | Needed for keratoconus screening and irregular corneas |
| Technique | How | Best for |
|---|---|---|
| Diffuse illumination | Wide beam with a diffuser, low magnification | General overview of lids, conjunctiva, cornea; contact lens fit and position |
| Direct focal (optic section and parallelepiped) | Narrow (optic section) or 1–2 mm wide beam focused on the tissue, light at 30–60° | Depth of corneal lesions, anterior chamber depth (van Herick), lens opacities |
| Conical beam | Small round spot, dark room, high magnification | Cells and flare in the anterior chamber (Tyndall effect) |
| Indirect illumination | Beam focused beside the area to be seen | Microcysts, epithelial oedema, iris sphincter |
| Retro-illumination | Light reflected from the iris or fundus behind the structure | Corneal vessels, keratic precipitates, lens vacuoles and posterior capsule opacity, iris transillumination |
| Specular reflection | Angle of illumination equals angle of viewing (about 60° apart), high magnification | Corneal endothelium (cell mosaic, guttae), tear film lipid layer |
| Sclerotic scatter | Beam at the limbus; light travels through the cornea by total internal reflection | Faint central corneal haze and oedema (central corneal clouding in contact lens wear) |
| Tangential | Light almost parallel to the surface | Iris surface, lens capsule elevations |
| Cobalt blue with yellow filter | Fluorescein, blue light, yellow barrier filter | Staining, tear break-up time, RGP fit, Seidel test |
Work in a logical order (lids, lashes, tear film, conjunctiva, cornea, anterior chamber, iris, lens), start with low magnification and diffuse light, and use a red-free (green) filter to see vessels and rose bengal or lissamine green staining.
Practical pearls
Where students slip
Related: Slit-Lamp Examination: Systematic Technique and Grading Scales, Corneal Topography in Ahmedabad, Topography and Keratoconus Screening Card: Cut-offs and Staging, Specular Microscopy Reference Card: Cell Density, CV and Hexagonality, Contact Lens Calculator: Soft, Toric and RGP Power, Base Curve and Tear Lens, Keratoconus: Signs, Indices, Staging, Progression and Treatment. Back to the optometry curriculum.
5,885 original MCQs, 150 OSCE stations and 800 viva questions for optometry licensing exams (NBEO, OCANZ, OEBC, UK and India), every option explained with a "why not" and a pearl.
How to cite this page
Patel D. Keratometry and Slit Lamp Illumination Techniques for Optometrists. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/keratometry-slit-lamp-techniques.html. Accessed [date].