Keratometry and Slit Lamp Illumination Techniques for Optometrists

Keratometry and Slit Lamp Illumination Techniques for Optometrists

Last updated: November 2026

RPC Rounds for Optometrists · B.Optom study guide

Keratometry and the slit lamp are the two instruments an optometrist uses most after the phoropter. Keratometry gives corneal curvature for contact lenses and IOL calculations; the slit lamp, used with the right illumination, shows everything from contact lens fit to early keratoconus.

Keratometry: principle

  • The cornea acts as a convex mirror. The keratometer measures the size of the image (mires) it reflects and calculates the radius: r = 2 × distance × image size ÷ object size.
  • Doubling (by prisms) splits the image so that eye movements do not disturb the reading; you align the doubled images rather than measure a moving one.
  • The result is converted to dioptres with the keratometric index 1.3375: K (D) = 337.5 ÷ r (mm). This fictitious index allows for the posterior corneal surface.
  • It measures only a small zone: about 3 mm across in the central cornea (the paracentral ring), not the apex.
InstrumentDesignNotes
Javal-SchiøtzFixed image size, variable object size; one-position (measures one meridian at a time)Mires are a red step and a green rectangle; overlap of steps gives astigmatism
Bausch and Lomb (Helmholtz type)Variable image size, fixed object; two-position (both meridians without rotating)Plus and minus signs on circles; align the pluses for horizontal and minuses for vertical
Automated keratometerMeasures reflected light points at several radiiFast; included in autorefractors and biometers
Topographer and tomographerPlacido or Scheimpflug imaging of the whole corneaNeeded for keratoconus screening and irregular corneas

Reading the mires

  • Regular astigmatism: the principal meridians are 90° apart; with-the-rule when the vertical meridian is steeper.
  • Irregular astigmatism (keratoconus, scars, post-graft): the mires are distorted or the meridians are not at 90°; refer for topography.
  • Normal range about 40–47 D (mean about 43–44 D). Readings over about 47 D, or a big difference between eyes, raise suspicion of keratoconus.
  • Range extension: a +1.25 D lens over the aperture extends the upper range (steep corneas), a −1.00 D lens the lower range.

Slit lamp illumination techniques

TechniqueHowBest for
Diffuse illuminationWide beam with a diffuser, low magnificationGeneral overview of lids, conjunctiva, cornea; contact lens fit and position
Direct focal (optic section and parallelepiped)Narrow (optic section) or 1–2 mm wide beam focused on the tissue, light at 30–60°Depth of corneal lesions, anterior chamber depth (van Herick), lens opacities
Conical beamSmall round spot, dark room, high magnificationCells and flare in the anterior chamber (Tyndall effect)
Indirect illuminationBeam focused beside the area to be seenMicrocysts, epithelial oedema, iris sphincter
Retro-illuminationLight reflected from the iris or fundus behind the structureCorneal vessels, keratic precipitates, lens vacuoles and posterior capsule opacity, iris transillumination
Specular reflectionAngle of illumination equals angle of viewing (about 60° apart), high magnificationCorneal endothelium (cell mosaic, guttae), tear film lipid layer
Sclerotic scatterBeam at the limbus; light travels through the cornea by total internal reflectionFaint central corneal haze and oedema (central corneal clouding in contact lens wear)
TangentialLight almost parallel to the surfaceIris surface, lens capsule elevations
Cobalt blue with yellow filterFluorescein, blue light, yellow barrier filterStaining, tear break-up time, RGP fit, Seidel test

Work in a logical order (lids, lashes, tear film, conjunctiva, cornea, anterior chamber, iris, lens), start with low magnification and diffuse light, and use a red-free (green) filter to see vessels and rose bengal or lissamine green staining.

Practical pearls

  • Ask the patient to blink just before a keratometry reading: a dry tear film distorts the mires.
  • Sclerotic scatter is the quickest way to see central corneal clouding in a long-term PMMA or low-Dk lens wearer.
  • Use specular reflection on every contact lens patient over 40 to check for guttae before fitting.

Where students slip

  • Forgetting that keratometry measures the paracentral cornea, not the apex.
  • Using keratometry alone to rule out keratoconus: early cones need topography or tomography.
  • Using too much light and magnification first and missing lid and lash disease.

Test yourself: flip cards

Quick quiz

Related: Corneal Topography in Ahmedabad, Topography and Keratoconus Screening Card: Cut-offs and Staging, Specular Microscopy Reference Card: Cell Density, CV and Hexagonality, Contact Lens Calculator: Soft, Toric and RGP Power, Base Curve and Tear Lens, Keratoconus: Signs, Indices, Staging, Progression and Treatment. Back to the optometry curriculum.

Key sources

  • Elliott DB. Clinical Procedures in Primary Eye Care. 5th ed. Elsevier; 2020
  • Henson DB. Optometric Instrumentation. 2nd ed. Butterworth-Heinemann; 1996
  • American Academy of Ophthalmology. BCSC Section 3: Clinical Optics (current edition), chapter on instruments

📘 Want more exam-ready material? See Most Commons in Ophthalmology on Kindle, or browse all my books.

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy

Consultant, Cataract & Refractive Surgery — among a small number of AIIMS-trained ophthalmologists practicing in Gujarat. Read full credentials & experience or view his 28 publications on ResearchGate.

📚 Also the author of OKAT Decoded, FRCOphth Decoded and other high-yield ophthalmology exam-prep books used by residents and PG aspirants.

Looking for an eye specialist in Ahmedabad? Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.