Ophthalmic Examination Lenses: 90D to 20D, Gonioscopy and Laser Lenses Compared

For: PG residents · Practising ophthalmologists

Last updated: November 2026

A good wide-field slit-lamp lens can handle most routine peripheral retina screening, with the indirect ophthalmoscope and scleral depression kept for high-risk eyes. This guide compares the lenses for each job, from 90D to 20D, gonioscopy and laser, with their pros, cons and current prices in India and the US.

Why look beyond the indirect ophthalmoscope

Indirect ophthalmoscopy remains the gold standard for the peripheral retina, but in a busy OPD it costs time and is hard on the examiner's neck and back. Slit-lamp lenses let you stay seated, use binocular stereo and adjustable magnification, and move straight to laser when you need to.

The guide starts with lenses for peripheral retina screening, then covers the rest of the kit: other slit-lamp lenses, indirect ophthalmoscopy (BIO) lenses, gonioscopy lenses and laser lenses.

At a glance: retina screening lenses

Wider field means lower magnification: the SuperQuad 160 gives the biggest overview and the Mainster Wide Field the sharpest detail among the contact lenses.

LensTypeField (static / dynamic)Image magLaser spot magBest for
Volk 90DNon-contact74° / 89°0.76×1.32×Small pupils, easy handling, posterior pole to mid-periphery
Volk SuperFieldNon-contact95° / 116°0.76×1.32×Everyday all-rounder: 90D magnification, wider field
Volk Digital Wide FieldNon-contact103° / 124°0.72×1.39×Wide, crisp pan-retinal scan; shorter working distance
Volk Super VitreoFundusNon-contact103° / 124°0.57×1.75×Widest non-contact view, small pupils (3–4 mm)
Volk SuperQuad 160Contact160° / 165°0.50×2.0×Maximum field: periphery screening, PRP, barrage
Volk QuadrAsphericContact120° / 144°0.51×1.97×Wide field through small pupils, PRP
Ocular Mainster Wide FieldContact118° / 127°0.68×1.50×Sharper detail to mid-periphery, PRP and focal laser
Goldmann three-mirrorContactCentral lens plus mirrors reaching the oraAbout 0.93–1.06× (varies by make)About 0.94–1.08×Ora, vitreous base, anterior breaks, detailed look at a lesion

Static vs dynamic field: static field is what you see with the eye and lens held still; dynamic field is the total you can reach by tilting the lens and moving your view around. Figures are the manufacturers' own.

Non-contact lenses

For routine dilated screening, a SuperField or Digital Wide Field adds about 30° of dynamic field over the 90D, with no anaesthesia, gel or patient discomfort. All of these lenses give an inverted, laterally reversed image.

Volk 90D

  • Pros: long working distance (about 7 mm), easy to manipulate, forgiving through small pupils.
  • Cons: narrowest field of the group (89° dynamic), so reaching the periphery takes a lot of steering.

Volk SuperField

  • Pros: same 0.76× magnification and 7 mm working distance as the 90D, with a 116° dynamic field. A strong choice for the one lens you pick up for every patient.
  • Cons: still needs good dilation and patient cooperation to approach the far periphery.

Volk Digital Wide Field

  • Pros: 124° dynamic field with high resolution and reduced reflections.
  • Cons: 4–5 mm working distance leaves less room to tilt and manipulate the lens; some examiners find it fiddly.

Volk Super VitreoFundus

  • Pros: 124° dynamic field designed for small pupils (3–4 mm), useful in diabetics, the elderly and pseudoexfoliation.
  • Cons: low magnification (0.57×), so fine detail is softer; short working distance (4–5 mm).

Shared limits: no scleral depression, a hand-held image that moves with every tremor, and a view of the ora that depends on full dilation and a cooperative patient. Non-contact lenses also collect oil from the lashes, so clean them often.

Wide-field contact lenses

Contact lenses rest on the cornea and move with the eye, so the view is steadier, the stereo is better and you can laser through the same lens. The price is topical anaesthesia, a coupling gel and a little patient discomfort.

Volk SuperQuad 160

  • Pros: the widest field of the slit-lamp lenses here (160° / 165°), from disc and macula out towards the ora. Excellent for spotting and treating peripheral tears, lattice and PRP in one sitting.
  • Cons: low magnification (0.50×), so small holes need a second look at higher slit-lamp magnification; the most expensive lens here.

Volk QuadrAspheric

  • Pros: 120° / 144° field and good performance through small pupils; flange, no-flange and ANF+ (advanced no fluid) versions are offered.
  • Cons: narrower than the SuperQuad 160 for a similar price.

Ocular Mainster Wide Field

  • Pros: crisper image (0.68×) with 1.50× laser spot magnification, good from posterior pole to mid-periphery; an NMR (no methylcellulose required) version exists.
  • Cons: 127° dynamic field, so the least peripheral reach of the wide-field contact lenses.

Goldmann three-mirror

  • Pros: upright image in the central lens with high magnification; the mirrors reach the ora and vitreous base more reliably than any wide-field lens. The best seated way to confirm a suspected break or retinal tuft.
  • Cons: slow (you scan mirror by mirror, clock hour by clock hour), so it suits a targeted look, not screening. Each mirror shows the retina opposite it: the mirror at 12 o'clock shows the inferior retina, upside down but not reversed side to side.

Shared limits: avoid with corneal epithelial defects, recent corneal surgery or infection; the gel blurs the cornea, so do OCT and photos first.

Other slit-lamp lenses: 60D to 78D and high magnification

Lower-power slit-lamp lenses give more magnification and better stereo but a smaller field, so they suit the disc and macula rather than the periphery. The rule of thumb holds across every series: as dioptric power rises, the field widens, magnification falls and working distance shortens.

LensField (static / dynamic)Image magLaser spot magWorking distanceBest forIndia (before GST)US list
Volk Digital High Mag57° / 70°1.30×0.77×13 mmDetailed look at a known disc or macular lesion₹37,900About $733
Volk Digital 1.0x60° / 72°1.0×1.0×12 mmDisc measurement without a correction factorAsk dealerAsk dealer
Volk 60D68° / 81°1.15×0.87×13 mmDisc and macula, glaucoma assessment₹31,226Ask dealer
Volk Super 6680° / 96°1.0×1.0×11 mmStereo disc and macula; disc size read directlyAsk dealerAbout $687–722
Volk 78D81° / 97°0.93×1.08×8 mmBalance of posterior pole detail and field₹31,226About $569–604
Volk SuperPupil XL103° / 124°0.45×2.2×4 mmVery small pupils (2–3 mm)Ask dealerAbout $710
  • 60D and Super 66: best stereo of the non-contact range. At 1.0× the Super 66 needs no conversion factor for disc size. Both need a well-dilated pupil.
  • 78D: better stereopsis than the 90D but usually needs dilation. A good one-lens choice for a glaucoma or medical retina clinic.
  • Disc size correction: to measure the vertical disc diameter with a thin slit beam, multiply the reading by the lens factor: about 1.3 for the 90D (1.39 in some studies), 1.1 for the 78D, 1.0 for the Super 66 and 0.88–1.0 for the 60D. The factors assume an average eye; high ametropia changes them.
  • High Mag lenses: for analysing a known abnormality, not for scanning.
  • Ocular Instruments MaxField series: equivalents of these lenses (66D, 78D, 84D, 90D, High Mag 78D) with larger quoted dynamic fields, for example 94° / 153° for the MaxField 90D. Ocular and Volk measure field differently, so compare them in your own hands.

Indirect ophthalmoscopy (BIO) lenses: 14D to 40D

The 20D and the Pan Retinal 2.2 are the all-round BIO lenses. The Pan Retinal 2.2 gives about 22% more field at a magnification close to the 20D's, with a working distance about 10 mm shorter. Use higher powers for small pupils and children, and lower powers for the disc and macula.

LensField (static / dynamic)Image magLaser spot magWorking distanceBest forIndia (before GST)US list
Volk Macula Plus 5.536° / 43°5.50×0.18×80 mmHighest-magnification view of the posterior poleAbout ₹32,200About $587
Volk 14D36° / 47°4.30×0.23×75 mmDisc and macula detailAsk dealerAbout $525
Volk 15D36° / 47°4.11×0.24×72 mmDisc and macula detailAsk dealerAbout $525
Volk Digital ClearMag38° / 49°3.89×0.26×60 mmDigital-series upgrade to the 14D / 15DAsk dealerAsk dealer
Volk 20D46° / 60°3.13×0.32×50 mmThe standard general diagnostic lens₹24,366About $497–532
Volk 25D52° / 68°2.54×0.39×38 mmBetween the 20D and 28DAsk dealerAbout $525
Volk Pan Retinal 2.256° / 73°2.68×0.37×40 mmWider field at near-20D magnification; small pupils₹24,366About $525–560
Volk Digital ClearField (now ClearField Pro)55° / 72°2.79×0.36×37 mmDigital-series upgrade to the 20D / Pan Retinal 2.2Ask dealerAbout $581 (Pro)
Volk 28D53° / 69°2.27×0.44×33 mmSmall pupils, fast scanningAsk dealerAbout $497–532
Volk 30D58° / 75°2.15×0.47×30 mmSmall pupils; slightly wider than the 28DAsk dealerAbout $525
Volk 40D69° / 90°1.67×0.60×20 mmWidest BIO field: small pupils, children, ROPAsk dealerAbout $525
  • 20D: best balance of field and magnification. The lens most of us learned on, and still the one to use with scleral depression.
  • Pan Retinal 2.2: a near-20D image with a field close to a 30D's. A worthwhile upgrade if you use the indirect only for selected patients and want each exam to be quicker.
  • 28D and 30D: wider field and shorter working distance, so easier through small pupils, but at lower magnification. Small holes are easier to miss.
  • 40D: low-magnification scanning out to the far periphery; mostly for paediatric work and ROP screening. Follow up anything suspicious with a higher-magnification lens.
  • 14D, 15D and Macula Plus 5.5: long working distance and high magnification for the disc and macula, for example in geriatric patients or glaucoma screening. The Macula Plus comes with an adapter to steady it at 80 mm.
  • Autoclavable (ACS) and single-use 20D and 28D versions are made for theatre and infection control.

Gonioscopy lenses

For routine gonioscopy, a four-mirror lens (Sussman, Posner or Zeiss style) is fastest and the only type that allows indentation. Keep a Goldmann-type lens for a stable view, laser and photography.

Every method needs a lens: light from the angle meets the tear film–air interface at more than the critical angle (about 46°), so it is totally internally reflected and the angle cannot be seen directly. The lens replaces that interface with one the light can cross.

LensTypeMirrors / designImage magCoupling gelIndia (before GST)US list
Volk G-1 (Goldmann single mirror)IndirectOne 62° mirror; rotate to see all quadrantsAsk dealerYes₹37,587About $716
Volk G-2 / Ocular two-mirrorIndirectTwo 62° mirrorsAsk dealerYes (NMR versions without)₹37,587 (G-2)About $716 (Volk G-2)
Volk G-3 / Goldmann three-mirrorIndirect, plus fundusMirrors at 60° (angle), 66° (equator to ora), 76° (mid-periphery)1.06× (flange), 1.03× (no flange)Yes with flange; no-flange versions need less₹37,587 (G-3); ₹30,018 (Volk three-mirror)About $716 (G-3); $568–637 (Volk three-mirror)
Volk G-4 (Sussman / Posner style)IndirectFour 64° mirrors; ring or Posner-style handle1.0×No₹37,587About $716
Volk G-4 High MagIndirectFour 64° mirrors1.50×No₹39,576About $733
Volk G-6IndirectSix mirrors, near-360° view without rotation1.0×No₹39,698About $724
Ocular Sussman four-mirrorIndirectFour 64° mirrors, hand-held, 9 mm contact0.80×NoAsk dealerAbout $460
Ocular KoeppeDirectDome-shaped lens resting on the cornea; patient supineAbout 1.5×Saline or gonio solutionAsk dealerAbout $440
Swan-Jacob, Hill, Volk Vold and other surgical gonioprismsDirect, intraoperativeUpright view for MIGS and goniotomyVariesViscoelasticAsk dealerAsk dealer

Goldmann-type (one-, two- and three-mirror)

  • Pros: steady, high-quality view; the best choice for angle laser and photography. The three-mirror also examines the fundus and ora.
  • Cons: needs gel, which blurs later fundus imaging; you must rotate it to see the whole angle; no indentation, so it cannot tell appositional from synechial closure.

Four-mirror (Sussman, Posner, Zeiss style)

  • Pros: no gel; all four mirrors face the angle, so you see 360° with little rotation; indentation (dynamic) gonioscopy separates appositional from synechial closure.
  • Cons: small contact area, harder to hold steady; slight pressure artificially opens the angle or causes corneal folds; less stable for laser.

Koeppe and other direct lenses

  • Pros: upright, panoramic view; lets you compare both eyes side by side; the standard for examination under anaesthesia and paediatric glaucoma.
  • Cons: patient must be supine, and you need a hand-held microscope or an operating microscope plus an illuminator.

Surgical gonioprisms (Swan-Jacob, Hill, Volk Vold, Mori upright) are used under the operating microscope for MIGS and goniotomy, not for diagnosis in clinic.

Reading the mirror: an indirect gonio lens shows the angle opposite the mirror. With the mirror at 12 o'clock you see the inferior angle; the image is flipped top to bottom but not side to side. Grading systems are covered in Gonioscopy Grading: Shaffer, Scheie and Spaeth.

Laser lenses

These contact lenses concentrate the beam and steady the eye. For a small anterior-segment practice, a Wise or Abraham lens for iridotomy plus a capsulotomy lens covers most YAG work.

LensUseImage magLaser spot magIndia (before GST)US list
Ocular Abraham Iridectomy (YAG / argon)Peripheral iridotomy; +66 D button1.5× / 1.6×0.67× / 0.63×Ask dealerAbout $625
Ocular Wise Iridotomy-SphincterotomyIridotomy, especially thick irides; +103 D button2.6×0.38×Ask dealerAbout $625
Volk Iridectomy / MagPlus IridectomyIridotomyAsk dealerAsk dealer₹41,314About $714
Volk CapsulotomyYAG posterior capsulotomy1.57×0.64×₹41,314About $722–875 (both prices listed on volk.com)
Ocular Abraham CapsulotomyYAG capsulotomy1.8×0.56×Ask dealerAbout $625
Volk SLT lensSLT; single mirror1.0×1.0×Ask dealerAbout $688
Volk Rapid SLTSLT; four simultaneous views of the meshwork, less rotationAsk dealerAsk dealerAsk dealerAbout $730–926 (both prices listed on volk.com)
Volk Area CentralisFocal/grid macular laser; 70° / 84° field1.06×0.94×₹64,923About $1,175
Ocular Mainster Standard Focal/GridFocal/grid laser; 90° / 121° field0.96×1.05×Ask dealerAbout $900
Ocular Mainster PRP 165PRP; 165° / 180° field0.51×1.96×Ask dealerAbout $900
  • Wise vs Abraham: the Wise's stronger button perforates the iris with less energy and shorter burns than the Abraham, and lowers energy density at the cornea and retina, according to Ocular's data sheet. The Abraham is gentler to learn on.
  • Laser spot magnification: above 1× the burn on the retina is larger than the setting, below 1× it is smaller. Always adjust the spot size for the lens: a 200 µm setting through a SuperQuad 160 (2×) gives about 400 µm on the retina, with four times the area. More detail is in Slit-Lamp Laser Lenses: Magnification and Spot Size Factors.
  • Single-use versions (Volk®1 capsulotomy, iridotomy, SLT) remove the need for disinfection, at a per-patient cost.

How far do we really see?

Measured from the centre of the eye, the ora serrata lies at roughly 220°, beyond the 200° that ultra-widefield cameras such as Optos are rated for. Anatomically it sits about 6–7 mm behind the limbus, a little more anterior nasally than temporally.

Field-of-view numbers are not comparable across devices, because manufacturers measure from different reference points:

  • Internal angle (from the eye's centre): used by Optos, which makes numbers look large. 200° covers roughly 80% of the retina.
  • External angle (from the pupil): used by Zeiss Clarus and most lens specifications. Clarus's 133° per image is in the same range as Optos's 200°.

So a "160°" lens is not 60° short of the ora. In practice a dynamic exam with a wide-field lens, with the patient looking into each quadrant, gets close to the ora in most eyes. Seeing the ora itself and the vitreous base still calls for a three-mirror lens or scleral depression.

Technique pearls

Full dilation and asking the patient to look towards the quadrant you are examining matter more than which lens you buy.

  1. Dilate fully. Most lenses here perform poorly below about 5 mm; the Super VitreoFundus and SuperPupil XL are the exceptions.
  2. Start at low slit-lamp magnification (10×) with a wide, moderately bright beam, focus on the disc, then move outward.
  3. Ask the patient to look towards the quadrant you want: look up for the superior retina, and so on.
  4. Tilt the lens slightly and shift the joystick the opposite way to push the view past the equator.
  5. Remember the image is inverted: a lesion at the top of your view is in the inferior retina.
  6. Contact lenses: topical anaesthetic, fill the cup with gel and no bubbles, have the patient look up, place the lower edge first, then tilt the lens onto the cornea.
  7. Afterwards, wipe off the gel and instil a lubricant drop to cut complaints of blurred vision.
  8. Document what you did, for example "wide-field contact lens examination, periphery to the equator flat, no breaks", rather than "periphery normal".

Cleaning and disinfection

  • Contact lenses (wide-field, three-mirror, gonio and laser lenses) touch the cornea and tear film, so they count as semi-critical devices: clean and then disinfect them between every patient, following the manufacturer's care guide for the agent and soak time.
  • Poorly disinfected contact lenses can spread adenovirus and other infections between patients.
  • Do not autoclave or heat-sterilise standard lenses; only the ACS (autoclavable) versions tolerate it.
  • Rinse off disinfectant fully before the next patient, and check the contact surface for scratches, which harbour debris and degrade the image.
  • Non-contact lenses: clean the surfaces with a lens cloth or the manufacturer's cleaner and keep them in their case; scratched coatings cause glare.

When indirect ophthalmoscopy is still a must

A wide-field lens can replace the indirect ophthalmoscope for low-risk routine screening, not for eyes at real risk of a retinal break. Do a dilated indirect examination with scleral depression, or at least a three-mirror examination, in these patients:

  • Acute flashes or floaters, new PVD, or vitreous haemorrhage. This is the standard of care, and a missed horseshoe tear here becomes a detachment.
  • High myopia, and before refractive, ICL or cataract surgery in eyes at risk.
  • Ocular trauma.
  • A personal history of retinal detachment in either eye, or a family history of it.
  • Stickler syndrome, Marfan syndrome and other high-risk conditions.
  • Any suspicious peripheral finding seen with a slit-lamp lens.

Scleral depression gives what no lens can: a dynamic view that shows vitreous traction and turns a flat-looking lesion into an obvious tear. See also the retinal detachment guide and fundus drawing colour codes.

Price comparison: retina screening lenses

In India, non-contact lenses cost about ₹31,000–35,000 and wide-field contact lenses about ₹69,000–76,000, including GST. Since 22 September 2025, GST on medical and ophthalmic instruments has been 5% (it was 12%). Some online listings still show the old 12% rate, so check the rate on your invoice.

LensIndia (before GST)India (incl. 5% GST, approx.)US list price
Volk 90D₹31,226₹32,790About $569–604
Volk SuperField₹29,392 (sale; MRP ₹35,270)₹30,860About $687–722
Volk Digital Wide FieldNot listed; ask a dealerNot listedAbout $733–768
Volk Super VitreoFundus₹33,439₹35,110About $710
Volk SuperQuad 160₹72,078₹75,680About $1,320
Volk QuadrAspheric₹65,732₹69,020About $1,221
Ocular Mainster Wide FieldNot listed; ask a dealerNot listedAbout $900
Volk three-mirror₹30,018₹31,520About $568–637

India prices are online listings from Medikabazaar before GST, with the total worked out at 5%. US prices are list prices on volk.com and ocularinc.com. All were checked in November 2026; list prices change often, and price ranges usually reflect ring colour or contact options.

Buying tips

  • Authorised distributors often quote 10–20% below online prices, especially for two or more lenses bought together; get a written quote.
  • Buy from authorised dealers so the warranty holds and to avoid counterfeit lenses.
  • Check the contact options before you buy (with flange, no flange, ANF+ or NMR); they change handling and whether you need gel.

Suggested kit and bottom line

A three-lens kit of a SuperField (or Digital Wide Field), a SuperQuad 160 and a three-mirror covers screening, laser and the ora for about ₹1.4 lakh including GST.

  1. Every dilated patient: SuperField or Digital Wide Field, with the Super VitreoFundus as the alternative if you see many small pupils.
  2. Steady wide view, or laser: SuperQuad 160 (or QuadrAspheric).
  3. Ora, vitreous base, confirming a break: three-mirror.
  4. High-risk and symptomatic eyes: indirect ophthalmoscope with scleral depression, still.

This split keeps the indirect ophthalmoscope for the small group that truly needs it, saving time and your back without lowering the standard of care.

🛒 Where to buy examination lenses

These are Amazon India search links and affiliate links: Sadbhaav Eye Clinic may earn a small commission from qualifying purchases at no extra cost to you. Check the seller, warranty and genuine-product guarantee before you buy; for costly devices, compare with authorised dealers.

Where residents slip

  • Calling the periphery "normal" after a 90D view that never passed the equator.
  • Forgetting the image is inverted and drawing a tear in the wrong quadrant.
  • Using the dial setting as the true spot size through a wide-field lens.
  • Grading the angle open through a four-mirror lens while pressing on the cornea.
  • Skipping scleral depression in a patient with acute PVD symptoms.

Related: Imaging devices compared, Indirect ophthalmoscopy and scleral depression, Slit-lamp laser lenses, Gonioscopy grading, Fundus examination for optometrists, SLT and laser iridotomy (patient guide), YAG capsulotomy (patient guide), Instruments Atlas.

Revise with flip cards

Key sources

📘 Want the exam-ready deep dive? This topic, and much more is covered in Ophthalmology Explorer (A High-Yield Clinical Reference for Ophthalmology Residents, Fellows & Board Exams) — available on Kindle. See all my books.

How to cite this page

Patel D. Ophthalmic Examination Lenses: 90D to 20D, Gonioscopy and Laser Lenses Compared. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/ophthalmic-examination-lenses.html. Accessed [date].

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy