Key points in 60 seconds
For a doctor in private practice, the theatre is your responsibility from the floor up: the autoclave, the indicators, the cleaning, the air and the records. This guide explains the classes of autoclave, the types of sterilisation indicators and how often to use each, and what fogging and fumigation can and cannot do. It is written for eye surgeons and other doctors setting up or running a small theatre, and for their OT staff.
The short answer
Steam kills microbes by transferring heat as it condenses on every surface. It only works if the steam touches every surface, so air must be removed from the chamber, the packs and the lumens of hollow instruments. Air pockets are the commonest reason sterilisation fails.
| Cycle | Temperature | Typical holding time | Approximate pressure |
|---|---|---|---|
| Standard (wrapped or unwrapped) | 121 °C | 15 to 20 minutes (longer for gravity cycles and large packs) | about 1.1 bar (15 psi) above atmospheric |
| Fast (pre-vacuum) | 134 °C | 3 to 4 minutes | about 2.1 bar (30 psi) above atmospheric |
| Extended (prion-risk instruments, where policy requires) | 134 °C | 18 minutes | about 2.1 bar |
Holding time is the time at full temperature, not the whole cycle. The whole cycle also includes air removal, heating, exhaust and drying. Always use the cycles your autoclave maker has validated.
The European standard EN 13060 sorts small steam sterilisers into three classes by what they can sterilise. Indian, Middle Eastern and many Asian suppliers use the same labels.
| Class | How air is removed | What it can sterilise | Pros | Cons |
|---|---|---|---|---|
| Class N ("naked") | Gravity: steam pushes air out | Solid, unwrapped instruments only | Cheapest; simple; short cycles | Cannot sterilise hollow, porous, wrapped or pouched items; instruments are not sterile once the door opens |
| Class S ("specified") | Single vacuum or other method; varies by model | Whatever the maker has tested and states, often some wrapped and some hollow items | Cheaper than Class B; some models cover simple pouched loads | You must read the maker's list of validated loads; not all S models handle lumens such as phaco tips |
| Class B ("big small") | Fractionated (repeated) pre-vacuum | All loads: wrapped, pouched, porous (drapes, gauze) and hollow (cannulas, handpieces) | Most reliable; dries loads; Bowie-Dick and Helix test cycles built in | Highest cost; longer cycles; needs good-quality water and regular servicing |
Other options you will meet:
| Check | Why it matters |
|---|---|
| Class B (or Class S validated for your exact loads) | Phaco handpieces, I/A cannulas, irrigating cannulas and pouched sets need vacuum air removal |
| Chamber size and number of trays | Enough for a full list without rushing short cycles; 18 to 23 litres suits most small eye theatres |
| Printer, USB or software log | A record of temperature, pressure and time for every cycle |
| Built-in Bowie-Dick, Helix and vacuum leak test programmes | Makes daily testing quick |
| Water quality requirement and supply | Most need distilled or reverse-osmosis water of low conductivity; hard water damages the chamber and instruments |
| Installation and performance validation (IQ, OQ, PQ) | Proves the machine works in your room with your loads; ask the supplier for documents |
| Local service engineer and spare parts | Seal and filter replacement, calibration and quick repair |
| Compatibility with your instruments | Check the handpiece and instrument makers' recommended cycles |
No single test proves a load is sterile. Use three kinds of monitoring together:
| Kind | What it shows | Pros | Cons |
|---|---|---|---|
| Physical (mechanical) | Gauges, display and printout of time, temperature and pressure | Every cycle; immediate | Shows chamber conditions only, not what happened inside a pack or lumen |
| Chemical | Ink or chemical changes colour when exposed to conditions | Cheap; immediate; can be used in and on every pack | Some types show only exposure, not sterility |
| Biological (BI) | Spores of Geobacillus stearothermophilus are killed or not | The only direct test of killing | Results take time (from under an hour with rapid-readout BIs to 24 to 48 hours); costs more; needs an incubator |
Chemical indicator types (ISO 11140-1). The type number tells you what the indicator measures; a higher number is not simply "better".
| Type | Example | What it tells you | Pros and cons |
|---|---|---|---|
| Type 1: process indicator | Autoclave tape, indicator printed on pouches | The pack went through a cycle | Separates processed from unprocessed packs; says nothing about sterility |
| Type 2: specific test | Bowie-Dick test pack or sheet; Helix PCD with strip | Air removal and steam penetration in pre-vacuum autoclaves | Essential daily test for Class B; not a load indicator |
| Type 3: single variable | Pellet that melts at a set temperature | One condition reached (usually temperature) | Rarely used now; ignores time and steam |
| Type 4: multi-variable | Strip that reacts to time and temperature | Two or more conditions reached | Better than types 1 and 3; still less exact than 5 and 6 |
| Type 5: integrating indicator | Moving-front strip | All critical conditions, matched to a biological indicator's performance | Good internal pack indicator; can be used to release non-implant loads where policy allows |
| Type 6: emulating indicator | Strip for one specific cycle (for example 134 °C, 3.5 minutes) | All conditions of that specific cycle | Very precise; must match the cycle exactly or results mislead |
These drawings show what a good and a bad result usually look like. Real products differ in colour and design, so keep the maker's colour chart next to the autoclave and train every staff member who releases loads.
If any indicator fails: do not use the pack; reprocess it; check how the load was packed and placed; run a Bowie-Dick or Helix test; and if the failure repeats, stop using the autoclave, call the engineer and recall items from earlier loads back to the last good biological indicator.
| When | What to do |
|---|---|
| Daily, before the first load | Warm-up cycle, then Bowie-Dick or Helix test; record the result |
| Every load | Check and keep the printout; type 1 indicator on every pack; type 4, 5 or 6 indicator inside each pack or tray; check dryness and indicator colour before release |
| At least weekly (many clinics do it daily) | Biological indicator in a process challenge device, with an unprocessed control BI from the same lot |
| Every load containing an implant | Biological indicator, and do not use the implant until the BI is negative (except in documented emergencies) |
| After installation, repair, relocation or a failed test | Three consecutive empty-chamber Bowie-Dick tests (pre-vacuum) and biological indicator tests before patient use |
| Every year | Service, calibration of temperature and pressure sensors, and requalification by the engineer |
| Date | Cycle no. | Programme | Load contents | Printout OK | Internal CI | BI result | Released by | Notes |
|---|---|---|---|---|---|---|---|---|
| 17/11/2026 | 0231 | 134 °C, 4 min, wrapped | Phaco set 1 and 2, cannula tray | Yes | Type 5 passed | Weekly BI negative at 24 h | Sister A | None |
Number each set and record which patient it was used for. If a biological indicator later turns positive, you can trace and recall every set from that load.
Fumigation fills a sealed room with a gas or vapour; fogging sprays a fine mist of liquid disinfectant. Both are meant to reach surfaces and air that cleaning misses. Neither replaces cleaning: dust and organic material shield microbes, so the room must be cleaned first.
| Method | How it works | Pros | Cons |
|---|---|---|---|
| Formaldehyde (formalin) fumigation | Formalin heated or mixed with potassium permanganate releases formaldehyde gas; neutralised with ammonia after | Cheap; long tradition; wide activity | Formaldehyde is a known human carcinogen and strong irritant; long closure (often overnight); residues; staff exposure. Indian public-health cleanliness (Kayakalp) guidelines advise replacing it with safer agents |
| Hydrogen peroxide with silver (for example 11% H2O2 with silver nitrate) | Fogging machine sprays a fine mist; dose per room volume | Breaks down to water and oxygen; shorter closure; widely used in Indian hospitals | Needs correct dose for the room volume and a closed room; can damage some electronics and metals; weaker on dirty surfaces |
| Hydrogen peroxide vapour or dry mist systems | Machine generates vapour or very fine particles of H2O2 | Strong evidence for killing spores on surfaces; good penetration | Expensive equipment; room must be sealed; trained operator |
| Ultraviolet-C (UV-C) devices | Mobile lamp irradiates surfaces | No chemicals; quick; no residue | Only works in line of sight; shadows; eye and skin hazard, so the room must be empty |
| Routine spraying of other disinfectants | Hand or machine spray of quaternary ammonium or alcohol products | Easy | Little evidence; uneven coverage; inhalation exposure |
What the evidence says. Clean surfaces, good air handling and good surgical practice prevent theatre infections. Fogging may help after construction work, after an outbreak or known contamination, or when reopening a room left closed for a long time. In theatres with a working HVAC system with HEPA filters and good cleaning, many infection-control bodies do not consider routine fogging necessary. If your accreditation body or hospital policy requires it, follow the policy and record it.
If you fog:
Never mix formalin with other chemicals except as a validated method, never enter a fumigated room without the stated wait and ventilation, and never fog with people, patients or food in the room. Staff handling formalin need gloves, goggles and a suitable respirator.
| When | What |
|---|---|
| Start of day | Damp-dust horizontal surfaces, lights and equipment; check temperature, humidity and pressure readings |
| Between cases | Clean and disinfect surfaces that touched the patient or were splashed (table, armrests, microscope handles, foot pedals, trolleys); remove waste and used linen |
| End of day (terminal clean) | Clean all surfaces and the floor wall-to-wall with detergent and disinfectant; clean scrub sinks; restock |
| Weekly | Walls, ceilings within reach, shelves, cupboards, air grilles; check and clean the autoclave chamber and water tank as the maker advises |
| Blood or body-fluid spills | Absorb, then disinfect with freshly made hypochlorite at your policy strength (often 1% available chlorine) or another approved product |
Is a pressure-cooker type autoclave acceptable for cataract surgery? Only for solid, unwrapped metal instruments with long enough cycles. For handpieces, cannulas and wrapped sets you need validated air removal: Class B, or Class S validated for those loads.
How long are sterilised packs good for? Most policies now use event-related shelf life: a pack stays sterile until something compromises it (tear, wetness, dropped, seal broken), provided it is stored clean, dry and off the floor. Some still use fixed dates; follow your policy and label every pack with the date and cycle number.
Do I need an incubator for biological indicators? Yes, a small BI incubator (or a rapid-readout reader) at the temperature the BI maker specifies, usually 55 to 60 °C for steam BIs.
Should I fog after every list? No. Clean between cases and do a terminal clean at the end of the day. Fog only when your policy requires it or for the special situations above.
See also: eye clinic infection control checklist and dispensing medicines at your clinic.
📘 Want more exam-ready material? See Most Commons in Ophthalmology on Kindle, or browse all my books.
How to cite this page
Patel D. Autoclaves, Sterilisation Indicators and OT Fogging: Class N, S and B Autoclaves, Bowie-Dick and Biological Indicators, Fumigation. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/autoclave-classes-sterilisation-indicators-ot-fogging.html. Accessed [date].