Set a timer for 40 minutes. The option order is mixed, so read every choice carefully.
1. Slit lamp illumination used to see corneal opacities against the red reflex is:
Answer: Retro-illumination. Light reflected from the iris or fundus back-lights the cornea.
2. Specular reflection on the slit lamp is used to view the:
Answer: Corneal endothelium. The endothelial mosaic is seen in the zone of specular reflection.
3. The van Herick technique estimates:
Answer: Peripheral anterior chamber depth. Compare peripheral chamber depth with corneal thickness at the limbus.
4. Gonioscopy is needed because the angle cannot be seen directly due to:
Answer: Total internal reflection at the cornea. A goniolens overcomes total internal reflection.
5. The indirect ophthalmoscope gives an image that is:
Answer: Inverted and real. Wide field, about 2–3× magnification with a 20 D lens.
6. The direct ophthalmoscope magnifies about:
Answer: 15×. Narrow field, upright image.
7. NCT (air-puff) tonometry is useful for:
Answer: Screening, without anaesthetic. Confirm high or unusual readings with Goldmann.
8. Rebound tonometry (iCare) is popular in children because:
Answer: It needs no anaesthetic and is quick. Light probe touches the cornea briefly.
9. The Schiotz tonometer is an:
Answer: Indentation tonometer. It uses weights and conversion tables; affected by scleral rigidity.
10. Before Goldmann tonometry, the prism should be:
Answer: Disinfected and dried as per protocol. Prevents transmission of infection.
11. Optical biometry (e.g. IOLMaster) may fail in:
Answer: Dense or white cataract. Use ultrasound A-scan when the optical signal is poor.
12. In contact A-scan, pressing too hard on the cornea gives an axial length that is:
Answer: Too short. Corneal compression shortens the measured axial length.
13. Keratometry values (K) are needed for IOL calculation because they give the:
Answer: Corneal power. Errors in K readings also cause refractive surprise.
14. A good A-scan of a phakic eye shows spikes from:
Answer: Cornea, anterior and posterior lens surfaces, retina and sclera. Retinal spike should be tall and steeply rising.
15. The HFA 10-2 programme is used to test:
Answer: Central 10°, e.g. advanced glaucoma and hydroxychloroquine screening. It tests points within 10° of fixation.
16. High false positives on a visual field mean the patient:
Answer: Pressed the button when no stimulus was shown. Trigger-happy patients; results look better than reality.
17. Mean deviation (MD) on a field printout indicates:
Answer: Overall average depression of the field. PSD shows localised irregularity.
18. OCT RNFL thickness is used to monitor:
Answer: Glaucoma. Thinning shows nerve fibre loss.
19. OCT of the macula is the key test for:
Answer: Diabetic macular oedema and wet AMD treatment decisions. It measures central thickness and fluid.
20. Fundus fluorescein angiography uses dye injected into a:
Answer: Vein, usually in the arm. Photographs are taken as dye passes through retinal vessels.
21. After FFA, patients should be told their urine and skin may:
Answer: Turn yellow-orange for a day or so. Harmless discolouration.
22. B-scan ultrasound is used when:
Answer: The view of the retina is blocked by dense cataract or haemorrhage. It shows retinal detachment, masses and vitreous haemorrhage.
23. Corneal topography is essential before:
Answer: Refractive surgery and to detect keratoconus. It maps the curvature of the whole cornea.
24. Pachymetry measures:
Answer: Corneal thickness. Ultrasound or optical; needed for glaucoma and refractive surgery.
25. Specular microscopy before cataract surgery is useful in:
Answer: Fuchs dystrophy or previous corneal surgery. Low cell counts raise the risk of corneal decompensation.
26. Schirmer test without anaesthetic measures:
Answer: Basal plus reflex tear secretion. Wetting under about 5 mm in 5 minutes suggests severe deficiency.
27. Tear break-up time under about 10 seconds suggests:
Answer: Unstable tear film. Measured with fluorescein after a blink.
28. Exophthalmometry (Hertel) measures:
Answer: Forward protrusion of the eye. Useful in thyroid eye disease.
29. The cover–uncover test detects:
Answer: Latent and manifest squint (tropia and phoria). The alternate cover test shows the total deviation.
30. The Worth four-dot test assesses:
Answer: Fusion and suppression. Uses red–green glasses.
31. Stereopsis can be tested with:
Answer: TNO or Titmus fly test. Measures depth perception in seconds of arc.
32. Contrast sensitivity is tested with:
Answer: Pelli-Robson chart. Patients can have good acuity but poor contrast, e.g. in cataract.
33. An autorefractor reading should be:
Answer: Refined with subjective refraction. Machines overestimate myopia in young people who accommodate.
34. Fundus photography is used to:
Answer: Document and compare retinal findings over time. Also used in diabetic screening programmes.
35. Calibration of a Goldmann tonometer is checked with a:
Answer: Calibration bar (weights) at 0, 2 and 6 positions. Check regularly and log the result.
36. Retinal photography with a non-mydriatic camera means:
Answer: Images can often be taken without dilating drops. Small pupils may still need dilation.
37. An ERG (electroretinogram) tests the function of the:
Answer: Retina. Used in inherited retinal disease.
38. Visual evoked potential (VEP) tests the:
Answer: Visual pathway up to the cortex. Delayed in optic neuritis.
39. The Placido disc assesses:
Answer: Corneal surface regularity. Distorted rings suggest irregular astigmatism.
40. Before an OCT scan, the most important patient instruction is to:
Answer: Keep looking at the fixation target and blink between scans. Movement and dry cornea reduce signal quality.
See the full curriculum to revise the topics behind these questions.
5,885 original MCQs, 150 OSCE stations and 800 viva questions for optometry licensing exams (NBEO, OCANZ, OEBC, UK and India), every option explained with a "why not" and a pearl.
How to cite this page
Patel D. Ophthalmic Technician Mock Test 4: Ophthalmic Instruments and Investigations (40 MCQs with Answers). RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/ophthalmic-technician-mock-test-4.html. Accessed [date].
Want to know more about treatment? Read about the Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.