Ophthalmic Technician Mock Test 4: Ophthalmic Instruments and Investigations (40 MCQs with Answers)

For: Allied eye-care staff

Last updated: November 2026

Key points in 60 seconds

  • 40 original questions on slit lamp illuminations, ophthalmoscopes, tonometers, biometry, perimetry, OCT, angiography, B-scan, topography, pachymetry, specular microscopy and tear tests.
  • Answer all of them, then press "Check my score": correct answers turn green and wrong ones red.
  • Open "Show answer" for the explanation.
  • Then try the other mock tests in the series.

Ophthalmic technician and eye OT assistant series: full curriculum · exams in India and worldwide · career and courses · study plan · mock tests 1, 2, 3, 4, 5 · eye OT assistant guide · practical and viva

Set a timer for 40 minutes. The option order is mixed, so read every choice carefully.

1. Slit lamp illumination used to see corneal opacities against the red reflex is:

Show answer

Answer: Retro-illumination. Light reflected from the iris or fundus back-lights the cornea.

2. Specular reflection on the slit lamp is used to view the:

Show answer

Answer: Corneal endothelium. The endothelial mosaic is seen in the zone of specular reflection.

3. The van Herick technique estimates:

Show answer

Answer: Peripheral anterior chamber depth. Compare peripheral chamber depth with corneal thickness at the limbus.

4. Gonioscopy is needed because the angle cannot be seen directly due to:

Show answer

Answer: Total internal reflection at the cornea. A goniolens overcomes total internal reflection.

5. The indirect ophthalmoscope gives an image that is:

Show answer

Answer: Inverted and real. Wide field, about 2–3× magnification with a 20 D lens.

6. The direct ophthalmoscope magnifies about:

Show answer

Answer: 15×. Narrow field, upright image.

7. NCT (air-puff) tonometry is useful for:

Show answer

Answer: Screening, without anaesthetic. Confirm high or unusual readings with Goldmann.

8. Rebound tonometry (iCare) is popular in children because:

Show answer

Answer: It needs no anaesthetic and is quick. Light probe touches the cornea briefly.

9. The Schiotz tonometer is an:

Show answer

Answer: Indentation tonometer. It uses weights and conversion tables; affected by scleral rigidity.

10. Before Goldmann tonometry, the prism should be:

Show answer

Answer: Disinfected and dried as per protocol. Prevents transmission of infection.

11. Optical biometry (e.g. IOLMaster) may fail in:

Show answer

Answer: Dense or white cataract. Use ultrasound A-scan when the optical signal is poor.

12. In contact A-scan, pressing too hard on the cornea gives an axial length that is:

Show answer

Answer: Too short. Corneal compression shortens the measured axial length.

13. Keratometry values (K) are needed for IOL calculation because they give the:

Show answer

Answer: Corneal power. Errors in K readings also cause refractive surprise.

14. A good A-scan of a phakic eye shows spikes from:

Show answer

Answer: Cornea, anterior and posterior lens surfaces, retina and sclera. Retinal spike should be tall and steeply rising.

15. The HFA 10-2 programme is used to test:

Show answer

Answer: Central 10°, e.g. advanced glaucoma and hydroxychloroquine screening. It tests points within 10° of fixation.

16. High false positives on a visual field mean the patient:

Show answer

Answer: Pressed the button when no stimulus was shown. Trigger-happy patients; results look better than reality.

17. Mean deviation (MD) on a field printout indicates:

Show answer

Answer: Overall average depression of the field. PSD shows localised irregularity.

18. OCT RNFL thickness is used to monitor:

Show answer

Answer: Glaucoma. Thinning shows nerve fibre loss.

19. OCT of the macula is the key test for:

Show answer

Answer: Diabetic macular oedema and wet AMD treatment decisions. It measures central thickness and fluid.

20. Fundus fluorescein angiography uses dye injected into a:

Show answer

Answer: Vein, usually in the arm. Photographs are taken as dye passes through retinal vessels.

21. After FFA, patients should be told their urine and skin may:

Show answer

Answer: Turn yellow-orange for a day or so. Harmless discolouration.

22. B-scan ultrasound is used when:

Show answer

Answer: The view of the retina is blocked by dense cataract or haemorrhage. It shows retinal detachment, masses and vitreous haemorrhage.

23. Corneal topography is essential before:

Show answer

Answer: Refractive surgery and to detect keratoconus. It maps the curvature of the whole cornea.

24. Pachymetry measures:

Show answer

Answer: Corneal thickness. Ultrasound or optical; needed for glaucoma and refractive surgery.

25. Specular microscopy before cataract surgery is useful in:

Show answer

Answer: Fuchs dystrophy or previous corneal surgery. Low cell counts raise the risk of corneal decompensation.

26. Schirmer test without anaesthetic measures:

Show answer

Answer: Basal plus reflex tear secretion. Wetting under about 5 mm in 5 minutes suggests severe deficiency.

27. Tear break-up time under about 10 seconds suggests:

Show answer

Answer: Unstable tear film. Measured with fluorescein after a blink.

28. Exophthalmometry (Hertel) measures:

Show answer

Answer: Forward protrusion of the eye. Useful in thyroid eye disease.

29. The cover–uncover test detects:

Show answer

Answer: Latent and manifest squint (tropia and phoria). The alternate cover test shows the total deviation.

30. The Worth four-dot test assesses:

Show answer

Answer: Fusion and suppression. Uses red–green glasses.

31. Stereopsis can be tested with:

Show answer

Answer: TNO or Titmus fly test. Measures depth perception in seconds of arc.

32. Contrast sensitivity is tested with:

Show answer

Answer: Pelli-Robson chart. Patients can have good acuity but poor contrast, e.g. in cataract.

33. An autorefractor reading should be:

Show answer

Answer: Refined with subjective refraction. Machines overestimate myopia in young people who accommodate.

34. Fundus photography is used to:

Show answer

Answer: Document and compare retinal findings over time. Also used in diabetic screening programmes.

35. Calibration of a Goldmann tonometer is checked with a:

Show answer

Answer: Calibration bar (weights) at 0, 2 and 6 positions. Check regularly and log the result.

36. Retinal photography with a non-mydriatic camera means:

Show answer

Answer: Images can often be taken without dilating drops. Small pupils may still need dilation.

37. An ERG (electroretinogram) tests the function of the:

Show answer

Answer: Retina. Used in inherited retinal disease.

38. Visual evoked potential (VEP) tests the:

Show answer

Answer: Visual pathway up to the cortex. Delayed in optic neuritis.

39. The Placido disc assesses:

Show answer

Answer: Corneal surface regularity. Distorted rings suggest irregular astigmatism.

40. Before an OCT scan, the most important patient instruction is to:

Show answer

Answer: Keep looking at the fixation target and blink between scans. Movement and dry cornea reduce signal quality.

See the full curriculum to revise the topics behind these questions.

Revise with flip cards

Key sources

  • Original questions written for this site; facts checked against standard ophthalmology, optics and pharmacology texts
  • Bio-Medical Waste Management Rules, 2016 (India)
  • WHO Surgical Safety Checklist
OptomDecoded: Comprehensive Optometry book cover

New for optometrists: OptomDecoded

5,885 original MCQs, 150 OSCE stations and 800 viva questions for optometry licensing exams (NBEO, OCANZ, OEBC, UK and India), every option explained with a "why not" and a pearl.

How to cite this page

Patel D. Ophthalmic Technician Mock Test 4: Ophthalmic Instruments and Investigations (40 MCQs with Answers). RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/ophthalmic-technician-mock-test-4.html. Accessed [date].

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy

Consultant, Cataract & Refractive Surgery — among a small number of AIIMS-trained ophthalmologists practising in Gujarat. Read full credentials & experience or view his 28 publications on ResearchGate.

📚 Also the author of OptomDecoded (optometry licensing exams), OKAT Decoded, FRCOphth Decoded, FICO Decoded, OKAP & ABO Decoded, FEBO Decoded and other high-yield ophthalmology exam-prep books used by residents and PG aspirants.

Want to know more about treatment? Read about the Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.