Phaco Machine Settings for Beginners: Peristaltic vs Venturi, Vacuum, Flow, Power and Fixing Surge

Phaco Machine Settings for Beginners: Peristaltic vs Venturi, Vacuum, Flow, Power and Fixing Surge

Last updated: November 2026

The phaco machine balances fluid coming in (inflow) with fluid and lens material going out (outflow) while ultrasound breaks up the nucleus. Understanding a few settings (vacuum, aspiration flow, inflow pressure and power) lets you set up any machine safely, recognise surge, and adjust when things are not working. Every machine is different, so use these as principles and your unit's presets as the starting point.

Inflow and outflow

  • Inflow comes from the bottle by gravity (bottle height, in cm) or from active pressurised systems that set the intraocular pressure in mmHg (e.g. Alcon Centurion active fluidics). Higher inflow keeps the chamber deeper and more stable.
  • Outflow is through the aspiration line (controlled by the pump) and through wound leak.
  • The chamber is stable when inflow is greater than or equal to outflow. When a block clears suddenly and outflow jumps, the chamber collapses: this is surge.

Peristaltic vs venturi pumps

FeaturePeristaltic (flow-based)Venturi (vacuum-based)
How it worksRollers squeeze the tubing; you set aspiration flow (mL/min) and a vacuum limitGas creates vacuum directly; you set vacuum; flow follows
Vacuum build-upOnly when the tip is occludedImmediate, even without occlusion
FeelControlled, good for beginnersFast attraction, strong followability
ExamplesMost machines including Alcon Infiniti/Centurion and many Indian machinesVenturi mode on Bausch & Lomb Stellaris, some others offer both

Typical settings by step (starting points only)

StepPowerVacuumAspiration flowComment
Sculpting (divide and conquer grooves)Moderate, continuous or pulseLow (about 50–100 mmHg)Low to moderate (about 20–25 mL/min)Low vacuum so the nucleus is shaved, not grabbed
Quadrant / chop removalPulse or burst, lowerHigh (about 250–450 mmHg)Moderate (about 30–40 mL/min)High vacuum to hold and bring fragments to the tip
EpinucleusLow or noneModerate (about 150–250 mmHg)ModerateAvoid catching the posterior capsule
Cortex (I/A)NoneHigh (about 400–600 mmHg) with a 0.3 mm portAbout 30–40 mL/minEngage cortex under the rhexis, pull centrally
Capsule polishNoneVery low (about 5–20 mmHg)LowGentle

Bottle height or set IOP is usually higher for chopping (e.g. 90–110 cm or the equivalent) and can be lowered for I/A. Values differ widely between machines and tips; your consultant's presets are the reference.

Ultrasound modes

ModeDescriptionUse
ContinuousConstant power while pedal is in position 3Sculpting
PulseFixed number of pulses per second with off-timeQuadrant removal; less heat
BurstSingle or repeated bursts of fixed durationChop, impaling the nucleus
Torsional / transversalSide-to-side tip movement (e.g. OZil)Better followability, less repulsion and heat

Watch for wound burn: high power with poor flow (occluded tip, tight wound, kinked sleeve, viscoelastic plugging the tip) heats the incision and can cause a corneal burn. Keep flow going and use pulse or torsional modes.

Before you start: priming and tune

Set-up checks

  1. Choose the right pack, tip (e.g. 0.9 mm, 30° or 45° bevel) and sleeve for your incision size.
  2. Prime the tubing and check for kinks, bubbles and leaks.
  3. Tune (test) the handpiece in the test chamber; fill the chamber completely.
  4. Check the bottle is full and at the right height (or IOP set), and the drip chamber is visible.
  5. Confirm the foot pedal positions and the settings for each step with the nurse.

Problems and fixes

ProblemLikely causeFix
Surge (chamber collapses after occlusion break)Vacuum too high for the inflow, low bottle, wound leakLower vacuum, raise bottle or set IOP, check the wound, use a smaller tip or surge-reducing settings
Shallow, unstable chamber throughoutLow bottle, leaking wound, kinked inflow tube, empty bottleRaise bottle, check tubing and bottle, suture or reduce wound leak
Poor followability (fragments do not come to the tip)Low flow or vacuum, wrong bevel orientationIncrease flow or vacuum; turn the bevel towards the fragment
Fragments chatter or bounce awayToo much power, too little vacuumLower power, increase vacuum, use pulse or torsional
Occlusion alarm with no progressTip blockedClear the tip with reflux; check aspiration tubing
Wound burnHigh power with low flowStop; restore flow; use less power

Pearls

  • Peristaltic pumps build vacuum only on occlusion: safest for beginners.
  • Low vacuum for sculpting, high vacuum for chopping and quadrant removal.
  • If the chamber surges, lower the vacuum or raise the inflow before changing anything else.

Common mistakes

  • Using chopping vacuum while sculpting, which grabs the nucleus and stresses zonules.
  • Ignoring a leaking wound that causes chamber instability.
  • High continuous power with an occluded tip (wound burn).

Test yourself: flip cards

Quick quiz

Related: Phacoemulsification Step by Step: A Beginner's Guide, Ophthalmic Viscosurgical Devices (OVDs): Cohesive, Dispersive, Viscoadaptive, Brands, Soft-Shell Technique and Removal, Ophthalmic Blades, Knives and Trocars: Keratome Sizes, Crescent, Side-Port, MVR and Vitrectomy Gauges Explained, Cataract Pre-operative Risk Checklist: Spot the Difficult Case, IOL Injectors and Cartridges for Residents: Alcon A, B, C, D, Preloaded Systems, J&J Unfolder, Zeiss, Medicel and Indian Lenses.

Key sources

  • Devgan U. Phaco fluidics and phaco ultrasound power modulations. Ophthalmol Clin North Am. 2006
  • Seibel BS. Phacodynamics: Mastering the Tools and Techniques of Phacoemulsification Surgery. 4th ed. Slack; 2005
  • Ward MS, Georgescu D, Olson RJ. Effect of bottle height and aspiration rate on postocclusion surge. J Cataract Refract Surg. 2008

📘 Want more exam-ready material? See Most Commons in Ophthalmology on Kindle, or browse all my books.

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy

Consultant, Cataract & Refractive Surgery — among a small number of AIIMS-trained ophthalmologists practicing in Gujarat. Read full credentials & experience or view his 28 publications on ResearchGate.

📚 Also the author of OKAT Decoded, FRCOphth Decoded and other high-yield ophthalmology exam-prep books used by residents and PG aspirants.

Looking for an eye specialist in Ahmedabad? Cataract surgery at Sadbhaav or book a consultation with Dr. Dhaval Patel.