Goldmann applanation (GAT) is still the reference standard, but its reading is affected by corneal thickness, curvature and rigidity.
Rebound tonometers (iCare) need no anaesthetic and suit children and screening, but agree with GAT only on average: do not swap between them for the same glaucoma patient.
Non-contact tonometers are screening tools; confirm raised readings with applanation.
Perkins and Tono-Pen are for patients who cannot sit at the slit lamp.
Dilute bleach is the disinfection method best supported for Goldmann prisms; alcohol wipes are linked to adenovirus outbreaks.
Every eye clinic measures pressure dozens of times a day, and the tonometer you pick changes both the number you write and how far you can trust it. This page compares the main tonometers, when to use each, how they agree with Goldmann, and how to keep them clean and calibrated.
At a glance
Tonometer
Principle
Anaesthetic
Best for
Main drawbacks
Goldmann applanation (GAT)
Imbert-Fick: force needed to flatten a 3.06 mm circle
Yes, with fluorescein
Reference standard for glaucoma diagnosis and follow-up
Needs a slit lamp and cooperation; affected by corneal thickness and astigmatism; prism disinfection
Perkins
Hand-held Goldmann-type applanation
Yes
Supine patients, children under sedation, theatre, wards
Technique-dependent; same corneal effects as GAT
Rebound (iCare IC100, IC200, PRO, HOME)
Speed at which a light probe rebounds off the cornea
No
Children, screening, home monitoring, patients who cannot tolerate GAT
Single-use probes; reads differently from GAT, more so at high pressures
Tono-Pen
Electronic applanation (Mackay-Marg principle)
Yes
Irregular or scarred corneas, bandage lenses, wards
Disposable covers; calibration drift
Non-contact (air puff)
Air pulse flattens the cornea; time or force is converted to pressure
No
High-volume screening by technicians
Less accurate at high pressures and in abnormal corneas; aerosol of tear film
Schiotz
Indentation by a known weight
Yes, supine
Where nothing else is available
Affected by scleral rigidity; largely obsolete
Dynamic contour (Pascal) and Ocular Response Analyzer
Contour matching; bidirectional air applanation with corneal hysteresis
Pascal yes; ORA no
Research, refractive surgery, glaucoma clinics interested in corneal biomechanics
Cost, availability
How well do they agree with Goldmann?
Rebound vs GAT: mean values agree well, but the limits of agreement are wide, so iCare and GAT readings are not interchangeable for following a glaucoma patient. The difference grows with central corneal thickness and age.
At high pressures rebound readings correlate less well with GAT; confirm any high or surprising reading with applanation.
In children rebound tonometers tend to read a little higher than applanation.
Thicker corneas make GAT, iCare and Tono-Pen read higher, and thin corneas (including after LASIK or PRK) make them read lower.
There is no validated formula to "correct" IOP for corneal thickness. Use pachymetry to interpret risk, not to adjust the number.
Getting a good Goldmann reading
Anaesthetic and fluorescein in the lower fornix; cobalt blue light at maximum brightness, beam wide open, at about 60° to the prism.
Set the dial at about 10 mmHg; ask the patient to look straight ahead and keep both eyes open; hold the lids against the orbital rim, not the globe.
Bring the prism onto the central cornea and turn the dial until the inner edges of the two semicircles just touch.
Mires that are too thick (too much fluorescein) read high; too thin read low.
With more than about 3 D of corneal astigmatism, rotate the prism so the red line on its holder is at the axis of the minus cylinder (about 43°), or average horizontal and vertical readings.
Record the time, because IOP varies through the day.
Disinfection and calibration
The AAO's technology assessment found that a dilute bleach (sodium hypochlorite) soak, about 1:10 for 5 minutes, reliably removed adenovirus and herpes simplex from prisms; 70% isopropyl alcohol reliably removed only herpes. Rinse well and dry before use.
Alcohol wipes and 3% hydrogen peroxide have been linked to adenovirus outbreaks. A 2025 analysis argued that alcohol wipes remain the most sustainable option, so practice still varies: follow your institution's policy.
All disinfectants damage prisms over time. Check prisms regularly for cracks and cloudiness, which can scratch the cornea and harbour microorganisms; replace damaged ones.
For suspected prion disease, use single-use tips or disposable covers: no disinfectant removes prions.
Every clinic: a slit-lamp-mounted Goldmann tonometer with a calibration bar and spare prisms.
Busy OPD with technicians: a non-contact tonometer for screening, with GAT for anyone above about 21 mmHg, glaucoma patients and suspects.
Children, home visits, camps: a rebound tonometer.
Theatre and wards: a Perkins or Tono-Pen.
Indian set-up guides list basic applanation or Schiotz tonometers at about ₹0.6–2 lakh, and a desktop non-contact tonometer is listed at about ₹3.5 lakh. Rebound tonometers need a dealer quote, and their single-use probes are a running cost. More in the equipment buying guide.
These are Amazon India search links and affiliate links: Sadbhaav Eye Clinic may earn a small commission from qualifying purchases at no extra cost to you. Check the seller, warranty and genuine-product guarantee before you buy; for costly devices, compare with authorised dealers.
Where residents slip
Pressing on the globe while holding the lids, which raises the reading.
Accepting thick mires from excess fluorescein.
Comparing this visit's iCare reading with last visit's Goldmann reading.
Wiping the prism with alcohol and putting it straight back on the next cornea.
📘 Want the exam-ready deep dive? This topic, and much more is covered in Ophthalmology Explorer (A High-Yield Clinical Reference for Ophthalmology Residents, Fellows & Board Exams) — available on Kindle. See all my books.
How to cite this page
Patel D. Tonometers Compared: Goldmann, Perkins, iCare, Tono-Pen and Non-Contact, With Disinfection. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/tonometers-compared.html. Accessed [date].
Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy