Tonometers Compared: Goldmann, Perkins, iCare, Tono-Pen and Non-Contact, With Disinfection

For: PG residents · Practising ophthalmologists

Last updated: November 2026

Key points in 60 seconds

  • Goldmann applanation (GAT) is still the reference standard, but its reading is affected by corneal thickness, curvature and rigidity.
  • Rebound tonometers (iCare) need no anaesthetic and suit children and screening, but agree with GAT only on average: do not swap between them for the same glaucoma patient.
  • Non-contact tonometers are screening tools; confirm raised readings with applanation.
  • Perkins and Tono-Pen are for patients who cannot sit at the slit lamp.
  • Dilute bleach is the disinfection method best supported for Goldmann prisms; alcohol wipes are linked to adenovirus outbreaks.

Every eye clinic measures pressure dozens of times a day, and the tonometer you pick changes both the number you write and how far you can trust it. This page compares the main tonometers, when to use each, how they agree with Goldmann, and how to keep them clean and calibrated.

At a glance

TonometerPrincipleAnaestheticBest forMain drawbacks
Goldmann applanation (GAT)Imbert-Fick: force needed to flatten a 3.06 mm circleYes, with fluoresceinReference standard for glaucoma diagnosis and follow-upNeeds a slit lamp and cooperation; affected by corneal thickness and astigmatism; prism disinfection
PerkinsHand-held Goldmann-type applanationYesSupine patients, children under sedation, theatre, wardsTechnique-dependent; same corneal effects as GAT
Rebound (iCare IC100, IC200, PRO, HOME)Speed at which a light probe rebounds off the corneaNoChildren, screening, home monitoring, patients who cannot tolerate GATSingle-use probes; reads differently from GAT, more so at high pressures
Tono-PenElectronic applanation (Mackay-Marg principle)YesIrregular or scarred corneas, bandage lenses, wardsDisposable covers; calibration drift
Non-contact (air puff)Air pulse flattens the cornea; time or force is converted to pressureNoHigh-volume screening by techniciansLess accurate at high pressures and in abnormal corneas; aerosol of tear film
SchiotzIndentation by a known weightYes, supineWhere nothing else is availableAffected by scleral rigidity; largely obsolete
Dynamic contour (Pascal) and Ocular Response AnalyzerContour matching; bidirectional air applanation with corneal hysteresisPascal yes; ORA noResearch, refractive surgery, glaucoma clinics interested in corneal biomechanicsCost, availability

How well do they agree with Goldmann?

  • Rebound vs GAT: mean values agree well, but the limits of agreement are wide, so iCare and GAT readings are not interchangeable for following a glaucoma patient. The difference grows with central corneal thickness and age.
  • At high pressures rebound readings correlate less well with GAT; confirm any high or surprising reading with applanation.
  • In children rebound tonometers tend to read a little higher than applanation.
  • Thicker corneas make GAT, iCare and Tono-Pen read higher, and thin corneas (including after LASIK or PRK) make them read lower.
  • There is no validated formula to "correct" IOP for corneal thickness. Use pachymetry to interpret risk, not to adjust the number.

Getting a good Goldmann reading

  1. Anaesthetic and fluorescein in the lower fornix; cobalt blue light at maximum brightness, beam wide open, at about 60° to the prism.
  2. Set the dial at about 10 mmHg; ask the patient to look straight ahead and keep both eyes open; hold the lids against the orbital rim, not the globe.
  3. Bring the prism onto the central cornea and turn the dial until the inner edges of the two semicircles just touch.
  4. Mires that are too thick (too much fluorescein) read high; too thin read low.
  5. With more than about 3 D of corneal astigmatism, rotate the prism so the red line on its holder is at the axis of the minus cylinder (about 43°), or average horizontal and vertical readings.
  6. Record the time, because IOP varies through the day.

Disinfection and calibration

  • The AAO's technology assessment found that a dilute bleach (sodium hypochlorite) soak, about 1:10 for 5 minutes, reliably removed adenovirus and herpes simplex from prisms; 70% isopropyl alcohol reliably removed only herpes. Rinse well and dry before use.
  • Alcohol wipes and 3% hydrogen peroxide have been linked to adenovirus outbreaks. A 2025 analysis argued that alcohol wipes remain the most sustainable option, so practice still varies: follow your institution's policy.
  • All disinfectants damage prisms over time. Check prisms regularly for cracks and cloudiness, which can scratch the cornea and harbour microorganisms; replace damaged ones.
  • For suspected prion disease, use single-use tips or disposable covers: no disinfectant removes prions.
  • Check Goldmann calibration with the calibration bar at least monthly. More detail is in the equipment calibration log and infection control checklist.

Which tonometer should a clinic buy?

  • Every clinic: a slit-lamp-mounted Goldmann tonometer with a calibration bar and spare prisms.
  • Busy OPD with technicians: a non-contact tonometer for screening, with GAT for anyone above about 21 mmHg, glaucoma patients and suspects.
  • Children, home visits, camps: a rebound tonometer.
  • Theatre and wards: a Perkins or Tono-Pen.

Indian set-up guides list basic applanation or Schiotz tonometers at about ₹0.6–2 lakh, and a desktop non-contact tonometer is listed at about ₹3.5 lakh. Rebound tonometers need a dealer quote, and their single-use probes are a running cost. More in the equipment buying guide.

🛒 Where to buy tonometers and supplies

These are Amazon India search links and affiliate links: Sadbhaav Eye Clinic may earn a small commission from qualifying purchases at no extra cost to you. Check the seller, warranty and genuine-product guarantee before you buy; for costly devices, compare with authorised dealers.

Where residents slip

  • Pressing on the globe while holding the lids, which raises the reading.
  • Accepting thick mires from excess fluorescein.
  • Comparing this visit's iCare reading with last visit's Goldmann reading.
  • Wiping the prism with alcohol and putting it straight back on the next cornea.

Related: Central corneal thickness and IOP, Tonometry for optometrists, Gonioscopy grading, Glaucoma medications chart, Visual field reference card.

Revise with flip cards

📘 Want the exam-ready deep dive? This topic, and much more is covered in Ophthalmology Explorer (A High-Yield Clinical Reference for Ophthalmology Residents, Fellows & Board Exams) — available on Kindle. See all my books.

How to cite this page

Patel D. Tonometers Compared: Goldmann, Perkins, iCare, Tono-Pen and Non-Contact, With Disinfection. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/tonometers-compared.html. Accessed [date].

Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy

Consultant, Cataract & Refractive Surgery — among a small number of AIIMS-trained ophthalmologists practising in Gujarat. Read full credentials & experience or view his 28 publications on ResearchGate.

📚 Also the author of OKAT Decoded, FRCOphth Decoded, FICO Decoded, OKAP & ABO Decoded, FEBO Decoded and other high-yield ophthalmology exam-prep books used by residents and PG aspirants.

Want to know more about treatment? Read about Glaucoma Treatment at Sadbhaav or book a consultation with Dr. Dhaval Patel.