RPC Rounds for Optometrists · B.Optom study guide
Measuring intraocular pressure (IOP) is a screening task optometrists do every day. A good optometrist knows which tonometer to use, how to get a reliable reading, what can make it wrong, and that a "normal" pressure does not rule out glaucoma.
| Tonometer | Principle | Strengths | Limitations |
|---|---|---|---|
| Goldmann applanation (GAT) | Force needed to flatten a 3.06 mm diameter area (Imbert-Fick law); fluorescein mires aligned at their inner edges | Reference standard; accurate | Needs slit lamp, anaesthetic and fluorescein; affected by corneal thickness and rigidity |
| Perkins | Hand-held Goldmann principle | Supine patients, children, theatre | Same corneal errors as GAT |
| Non-contact (air-puff, NCT) | Air pulse flattens the cornea; time or force to applanation converted to IOP | No anaesthetic, no contact, quick screening | Less accurate at high IOP; startles patients; take an average of 3 readings |
| Rebound (iCare) | Small probe bounces off the cornea; deceleration relates to IOP | No anaesthetic, portable; children and home monitoring | Reads slightly higher than GAT in some eyes; affected by CCT |
| Schiøtz (indentation) | Weight indents the cornea; scale reading converted with tables | Cheap, portable | Affected by ocular rigidity (high myopia, post-surgery); largely historical |
| Tono-Pen and pneumatonometer | Electronic applanation | Scarred or irregular corneas | Less accurate than GAT on normal corneas |
| Dynamic contour and ocular response analyser | Contour matching or corneal hysteresis | Less dependent on corneal properties | Specialist use |
| Factor | Effect |
|---|---|
| Thin cornea (or after LASIK/PRK) | Reads falsely low |
| Thick cornea | Reads falsely high (but oedematous corneas often read low) |
| Corneal astigmatism over 3 D | Error unless the prism is rotated to the minus-cylinder axis |
| Squeezing, tight collar, breath holding, pressure on the lids | Reads falsely high |
| Too thick or thin fluorescein mires | High or low readings |
| Repeated measurements | IOP falls slightly (tonography effect) |
There is no reliable formula to "correct" IOP for central corneal thickness; treat thin CCT as a risk factor. See CCT and IOP.
Practical pearls
Where students slip
Related: Eye Pressure Test (Tonometry), Central Corneal Thickness and IOP: Why There Is No Correction Formula, POAG and Ocular Hypertension: Diagnosis, Risk and Treatment, Glaucoma Notes for Residents: Topics and Syllabus, When to Refer: Eye Referral Urgency Guide for Optometrists. Back to the optometry curriculum.