For: Practising ophthalmologists
Last updated: November 2026
Key points in 60 seconds
Eye surgery depends on precise machines. A tonometer that reads 4 mmHg low, a biometer with a drifting axial length, or an autoclave that no longer pulls a vacuum can harm patients long before anyone notices. A simple schedule of checks and a logbook prevent most problems.
| Equipment | Daily or before use | Periodic | Notes |
|---|---|---|---|
| Goldmann applanation tonometer | Clean and disinfect the prism; check it is undamaged | Calibration check with the calibration bar at least monthly (at the 0, 20 and 60 mmHg positions); service if outside the maker's tolerance | Replace cracked prisms; consider disposable prisms |
| Non-contact and rebound tonometers | Self-test as per maker | Annual service and calibration | Rebound tonometer probes are single use |
| Optical biometer (IOLMaster, Lenstar and others) | Check the test eye if the maker recommends daily | Calibration check after moving, service or software updates; annual service | Compare readings between eyes and with previous data; see biometry checklist |
| A-scan ultrasound | Check probe and test block | Annual calibration | Use the correct velocity settings |
| Keratometer and topographer | Calibration sphere check as per maker | Annual service | Clean mires and placido cone |
| Auto-refractor, lensometer | Model eye or test lens check | Annual service | Lensometer: check zero and a known lens monthly |
| Slit lamp | Clean, check bulb and focus | Annual service; replace bulbs before failure | Keep a spare bulb |
| Operating microscope | Check illumination, foot pedal, focus and zoom; clean optics with the right solution | Annual service | Check X-Y and tilt, lamp hours |
| Phaco and vitrectomy machines | Tune and test (prime, fluidics, vacuum) before each list | Preventive maintenance every 6 to 12 months under a service contract | Log error codes |
| Lasers (YAG, green, SLT) | Aiming beam and test fire as per maker; check interlocks and warning signs | Annual service and power calibration | Laser safety officer, protective eyewear for staff |
| Autoclave | Bowie-Dick or Helix test; check printout | Leak test weekly; annual validation and calibration | See autoclaves and indicators |
| OT air conditioning and HEPA | Check temperature, humidity and pressure display | Filter change as scheduled; HEPA integrity and air-change validation yearly | Record pressure differential daily |
| Medicine fridge | Temperature log twice daily | Annual check of thermometer accuracy | Alarm for power cuts if possible |
| Electrical safety | Check for damaged cables | Earthing and leakage current test yearly; UPS battery check | Do not use multi-plug boards for theatre machines |
| Emergency equipment | AED self-test light, oxygen pressure, suction | Monthly full check | See crash cart checklist |
| Date | Equipment and serial no. | Check or event | Result | Action taken | By |
|---|---|---|---|---|---|
| 03/11/2026 | Goldmann tonometer, SN 1234 | Monthly calibration check | Within tolerance at 0, 20, 60 | None | Optometrist R |
| 05/11/2026 | Phaco machine, SN 5678 | Error code during tune | Vacuum sensor fault | Engineer called; loan machine used | OT in-charge |
| 10/11/2026 | Biometer, SN 9012 | Test eye check after software update | Passed | Recorded | Optometrist R |
📘 Want more exam-ready material? See Most Commons in Ophthalmology on Kindle, or browse all my books.
How to cite this page
Patel D. Ophthalmic Equipment Maintenance and Calibration: Daily Checks, Tonometer and Biometer Calibration, Service Contracts and a Log Template. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/ophthalmic-equipment-maintenance-calibration-log.html. Accessed [date].
Looking for an eye specialist in Ahmedabad? Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.