Short answer: small biometry errors cause big refractive surprises. A 1 mm error in axial length causes roughly 2.5–3 D of refractive error (more in short eyes), and a 1 D error in keratometry causes roughly 1 D of error. Check every result against the fellow eye and against expected ranges before choosing the IOL.
Error rules of thumb
Measurement error
Approximate refractive effect
Axial length 1 mm
About 2.5–3 D (larger in short eyes)
Axial length 0.1 mm
About 0.25–0.3 D
Keratometry 1 D
About 1 D
Wrong eye's data or wrong IOL
Unpredictable and potentially large; prevent it with a theatre check
Red flags: recheck the biometry if
Axial length differs between the two eyes by more than about 0.3 mm without an explanation (such as anisometropia).
Average keratometry differs between eyes by more than about 1 D.
Average K is very flat or very steep (for example, below about 40 D or above about 47 D).
Axial length is unusually short or long (for example, below about 22 mm or above about 25 mm).
The calculated IOL power doesn't fit the patient's refractive history.
The signal-to-noise ratio or quality indicators on the biometer are poor.
Optical vs ultrasound biometry
Optical biometry is more accurate and repeatable and is the standard where available. Dense or posterior subcapsular cataracts can block the measurement.
Contact A-scan compresses the cornea and tends to give a falsely short axial length (a myopic surprise). Immersion A-scan avoids compression and is preferred when ultrasound is needed.
Silicone oil-filled eyes: sound travels much more slowly in silicone oil (around 980 m/s), so ultrasound readings must be corrected. Optical biometers have a silicone oil mode.
Staphyloma in high myopia: measure along the visual axis (optical biometry, or B-scan-guided A-scan).
Keratometry pitfalls
Dry eye and MGD cause inconsistent K readings. Treat the surface first and repeat.
Contact lens wear changes corneal shape; ask the patient to stop lenses before measurement.
Use modern formulas (for example, Barrett Universal II, Kane, Hill-RBF or similar) rather than older regression formulas, especially in short or long eyes.
Optimise lens constants for your own results whenever you can.
Frequently asked questions
How much refractive error does a 1 mm axial length error cause?
About 2.5–3 D in an average eye, and more in short eyes. Even 0.1 mm causes about a quarter of a dioptre.
When should biometry be repeated?
When there is a large difference between the two eyes, unusual axial length or keratometry, poor quality readings, or a result that doesn't fit the patient's refractive history.
Why does contact A-scan give a myopic surprise?
The probe compresses the cornea, shortening the measured axial length. A shorter axial length leads to a stronger IOL, which leaves the patient myopic.
Patel D. Biometry Rules of Thumb: Axial Length, Keratometry and Red Flags. RPC Rounds (Sadbhaav Eye Clinic). Updated October 2026. Available at: https://www.sadbhaavclinic.com/biometry-rules-of-thumb.html. Accessed [date].