A one-page check before you sign an IOL power. Most refractive surprises come from a measurement that should have been repeated, not from the formula.
| Check | Recheck when |
|---|---|
| Axial length (AL) | Under about 22 mm or over about 25 mm, or more than 0.3 mm different from the fellow eye without anisometropia to explain it |
| Mean keratometry (K) | Under about 40 D or over about 47 D, or more than 1 D different between eyes |
| Corneal cylinder | Different axis or magnitude between the biometer and topography (repeat; treat dry eye first) |
| Calculated IOL power | More than about 3 D different between the two eyes, or not in keeping with the old spectacle prescription |
| Quality | Low signal-to-noise or standard deviation flags, a dense cataract, poor fixation |
| History | Previous LASIK, PRK or RK; previous vitrectomy or silicone oil; contact lens wear; keratoconus |
| Axial length | Good choices | Notes |
|---|---|---|
| Short (under 22 mm) | Barrett Universal II, Kane, Hill-RBF, Haigis; Hoffer Q among the older formulas | Check ACD and lens thickness; consider a nanophthalmos work-up if AL is under about 20.5 mm |
| Average (22–26 mm) | Barrett Universal II, Kane, EVO, Hill-RBF; SRK/T and Holladay 1 are acceptable | Use optimised constants |
| Long (over 26 mm) | Barrett Universal II, Kane, EVO; SRK/T or Holladay 1 with the Wang-Koch AL adjustment | Older formulas tend to leave long eyes hyperopic; aim slightly myopic |
| Post-refractive surgery | Barrett True-K, total-keratometry-based formulas, and the ASCRS calculator | See post-refractive IOL calculation |
| Situation | What to do |
|---|---|
| Silicone oil in the eye | Use the optical biometer's silicone oil mode. For ultrasound, sound velocity in oil is about 980 m/s (versus 1532 m/s in vitreous). Adjust the IOL power for the oil, typically adding about +3 to +3.5 D for a posterior-chamber IOL if the oil will stay. |
| Staphyloma or high myopia | Measure along the visual axis with optical biometry, or use B-scan-guided A-scan |
| Sulcus IOL (posterior capsule rupture) | Reduce the power, typically by about 0.5–1.0 D depending on the IOL power (see sulcus power adjustment). Use optic capture when possible. |
| Corneal astigmatism of about 1.0–1.25 D or more | Consider a toric IOL or limbal relaxing incisions, and account for posterior corneal astigmatism (use a toric calculator that includes it, such as Barrett). See toric IOL pearls and the LRI calculator. |
| Children | Aim for undercorrection (hyperopia) depending on age, because the eye will grow. See congenital cataract. |
Biggest preventable errors: the wrong eye's biometry, the wrong IOL taken from the shelf, and not repeating a measurement that "looked odd". A two-person check of eye, model and power prevents most of them.