Key points in 60 seconds
Ophthalmic Imaging course: Course home · M1 · M2 · M3 · M4 · M5 · M6 · M7 · M8 · M9 · M10
Practise: Mock 1 · Mock 2 · Mock 3 · Viva and OSCE · Career guide
Exam map: Optical coherence tomography: physics, time-domain vs spectral-domain vs swept-source, scan protocols, signal quality and artefacts.
| Type | Wavelength | Speed | Strengths |
|---|---|---|---|
| Time-domain (older) | ~820 nm | Slow (hundreds of A-scans/s) | Historical |
| Spectral-domain (SD-OCT) | ~840 nm | Tens of thousands of A-scans/s | High resolution retina; widely used |
| Swept-source (SS-OCT) | ~1,050 nm | 100,000+ A-scans/s | Deeper choroid, less affected by cataract, wide scans |
| Protocol | Use |
|---|---|
| Macular cube / volume | Thickness maps, fluid, follow-up of AMD, DME |
| High-resolution line or radial scans | Detail of holes, membranes, CNV |
| RNFL circle scan (about 3.4–3.5 mm diameter around the disc) | Glaucoma |
| Ganglion cell analysis | Early glaucoma, neuro-ophthalmology |
| ONH volume / BMO-MRW | Disc analysis |
| Anterior segment OCT | Angle, cornea, LASIK flaps, ICL vault |
| Artefact | Cause | What to do |
|---|---|---|
| Low signal strength | Dry cornea, cataract, small pupil, poor focus | Blink, lubricate, refocus, dilate |
| Segmentation error | Pathology, low signal, high myopia | Check B-scans; manual correction if allowed |
| Decentration | Poor fixation or alignment | Re-centre; use follow-up registration |
| Motion artefact | Eye movements, blinks | Tracking, repeat |
| Mirror artefact | Image near/past zero-delay line | Reposition |
| Floater shadows | Vitreous opacities | Ask patient to move eyes then rescan |
| Out-of-normative-database | High myopia, unusual discs, children, ethnicity | Interpret with caution |
Go deeper: retinal layers and OCT, imaging devices compared.
The bandwidth (coherence length) of the light source.
Its longer wavelength (~1,050 nm) penetrates deeper and scatters less.
Enhanced depth imaging: placing the zero-delay line near the choroid on SD-OCT.
Normative database comparisons are age-adjusted.
Dry cornea, cataract, small pupil (also poor focus, media opacity).
Answer all, then press “Check my score”. Option order is mixed.
1. OCT is based on:
Answer: Low-coherence interferometry. Light echo.
2. A B-scan in OCT is:
Answer: A cross-sectional image. Many A-scans.
3. SD-OCT commonly uses light around:
Answer: 840 nm. Near infrared.
4. RNFL circle scans are centred on the:
Answer: Optic disc. Glaucoma.
5. Segmentation errors are checked by:
Answer: Reviewing the B-scans. Always.
6. Eye tracking reduces:
Answer: Motion artefacts. Stability.
7. Normative databases may mislead in:
Answer: High myopes and children. Outside range.
8. Anterior segment OCT measures:
Answer: Angle and corneal thickness. Front of eye.
← M4. Anterior Segment, External and Gonio Photography | M6. OCT Interpretation: Macula, Glaucoma and Anterior Segment →
5,885 original MCQs, 150 OSCE stations and 800 viva questions for optometry licensing exams (NBEO, OCANZ, OEBC, UK and India), every option explained with a "why not" and a pearl.
How to cite this page
Patel D. Ophthalmic Imaging Module 5: OCT Principles, Scanning Technique and Artefacts. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/ophthalmic-imaging-m5-oct-principles-technique-artefacts.html. Accessed [date].
Want to know more about treatment? Read about the Eye Department at Sadbhaav or book a consultation with Dr. Dhaval Patel.