Five published formulas side by side: SRK/T, Holladay 1, Hoffer Q, Haigis and SRK II (teaching).
Shows the suggested IOL power for your target and the expected refraction for nearby powers.
Suggests a toric IOL when corneal astigmatism, after the incision, is at least 1.00 D with-the-rule or oblique, or 0.75 D against-the-rule.
Flags unusual biometry, differences between eyes, long eyes and previous refractive surgery; gives the sulcus power.
For learning and cross-checking: use Barrett, Kane or EVO online and your biometer for the final decision.
This calculator works out IOL power with five published formulas side by side: SRK/T, Holladay 1, Hoffer Q, Haigis and SRK II. It suggests the power closest to your target, shows the expected refraction for nearby powers, flags unusual biometry, gives the sulcus power, and tells you when a toric IOL should be preferred.
Use it to learn and to cross-check, not as your only calculation. Newer formulas such as Barrett Universal II, Kane, EVO and Hill-RBF are more accurate in most eyes, but their equations are not published, so they cannot be reproduced here. Use them on their official websites (links below) and in your biometer.
Enter biometry for the eye to be operated. Results update as you type. Nothing is stored or sent.
Toric planning: incision and surgically induced astigmatism
For a temporal incision in the right eye use 180°, left eye 0° (or your own meridian). Use your own measured SIA if you know it.
More options: other eye, age, long eyes, Haigis constants
For learning and cross-checking only. Base surgical decisions on your biometer's software and validated online calculators, with lens constants optimised for your IOL.
Exact power is the unrounded power for your target.
Suggested IOL is the available power (in 0.5 D steps) whose expected refraction is closest to the target, leaning slightly myopic when two are equally close.
Nearby powers show the expected spectacle refraction for 1.0 D either side, so you can choose for monovision or a second lens model.
A green tick marks the formulas that suit this axial length. If they disagree by more than 0.5 D, re-check the biometry before choosing.
The copy summary button gives a one-line note for your EMR or case sheet.
Which formula for which eye
Axial length
Formulas that perform well (among these five)
Notes
Below 22 mm (short)
Hoffer Q, Haigis
Small errors in ELP matter most; Barrett and Kane are preferred where available
22 to 24.5 mm (average)
SRK/T, Holladay 1, Hoffer Q, Haigis
All similar; optimise your lens constant
24.5 to 26 mm
SRK/T, Holladay 1, Haigis
Above 26 mm (long)
SRK/T, Haigis
Holladay 1 and Hoffer Q tend to leave hyperopia unless the Wang-Koch adjustment is applied; consider a slightly myopic target
After LASIK, PRK or RK
None of these
Use post-refractive methods (Barrett True-K, Haigis-L, Shammas) on the ASCRS calculator
A toric IOL is suggested when corneal astigmatism, after allowing for your incision, is 1.00 D or more with-the-rule or oblique, or 0.75 D or more against-the-rule. The lower against-the-rule threshold allows for posterior corneal astigmatism, which adds against-the-rule cylinder that front-surface keratometry misses.
Between 0.25 D below the threshold and the threshold, it suggests a borderline decision: a low-power toric or limbal relaxing incisions.
The IOL-plane cylinder is estimated as about 1.46 times the corneal cylinder. Always confirm the toric model and axis with the Barrett Toric or manufacturer's calculator, which include posterior corneal astigmatism and your actual IOL.
Check that keratometry from two devices agrees in magnitude and axis before ordering a toric lens (use the keratometry tools).
Lens constants
Enter the A-constant for SRK/T from the lens box, the manufacturer, or better, an optimised constant from IOLCon or your own outcomes. See the A-constant reference.
Holladay surgeon factor, Hoffer pACD and SRK/T ACD constant are converted from the A-constant using the standard Retzlaff and Holladay relationships.
For Haigis, enter optimised a0, a1 and a2 for your lens if you have them. Otherwise a0 is estimated from the A-constant (with a1 = 0.4 and a2 = 0.1) so that Haigis matches SRK/T in an average eye; this is an approximation.
Optical biometry constants differ from ultrasound (applanation) constants by about 0.3 to 0.4 in the A-constant.
The formulas in brief
Formula
Year and type
Predicts lens position from
SRK II
1988, regression
Axial length band (not used for planning today)
SRK/T
1990, theoretical with regression
Axial length and K (corneal height)
Holladay 1
1988, theoretical
Axial length and K, with surgeon factor
Hoffer Q
1993, theoretical
Axial length and K, with pACD
Haigis
1999, theoretical
Measured ACD and axial length, with three constants
Key sources
Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333-340 (and erratum)
Holladay JT, et al. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988;14:17-24
Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700-712; errata 1994 and 2007
Haigis W, et al. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765-773
Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011;37:2018-2027
Koch DD, et al. Contribution of posterior corneal astigmatism to total corneal astigmatism. J Cataract Refract Surg. 2012;38:2080-2087
Patel D. IOL Power Calculator: SRK/T, Holladay 1, Hoffer Q and Haigis Side by Side, with Toric IOL Suggestion. RPC Rounds (Sadbhaav Eye Clinic). Updated November 2026. Available at: https://www.sadbhaavclinic.com/iol-power-calculator-srkt-holladay-hoffer-haigis.html. Accessed [date].
Written and medically reviewed by Dr. Dhaval Patel, MD (Ophthalmology, AIIMS New Delhi) · Editorial policy